75
Method of reading density number
When flow unit is mL/*, up to 10 densities can be registered.
To register a new density, read the unconfigured density number
and then register according to the procedure of the setting method
of calculation time.
Continuing to hold down the PRINT key in weighing display
displays .
Blinking is the current density number and is the set density
value.
The density number can be changed by following key operation
The setting range is F01
~
F10.
RE-ZERO(+) key
Change density number.
MODE(-) key
Change density number.
PRINT key
Read the density of the selected density
number and return to weighing display.
CAL key
Return to the weighing display without
reading the density of the selected density number.
14-2 Flow Measurement Setting
Flow value range by models
Model
Flow value range
0.001g model 0.01~100
0.01g model
0.1~1000
0.1g model
1~5000
Depending on the sample to measure the flow rate, the above values may not apply.
Estimated flow rate value and calculating time (Ct) depending on your model
Calculation time automatic setting mode
By setting "Calculation time automatic mode" of Function table "Application Function
ap fnc
" to "
1
",
the optimal Ct is automatically set for the flow rate being measured. So, it is unnecessary to manually
set Ct.
.
.
Weighing display