79
16.Minimum Weighing Warning Function
The minimum weighing value is the minimum necessary amount of sample to be used for correctly
performing quantitatively performing quantitative analysis, taking into consideration measurement error of
the balance.
If the amount of sample is too small, the proportion of the measurement error in the measured value
increases accordingly, and the reliability of the analysis result may drop.
By using the minimum weighing warning function, it is possible to judge at a glance whether the amount of
sample meets the set minimum weight value. This function can be used only in "mg or g" mode.
" " is displayed at the top of the unit part when in use.
When the amount of sample is less than the set minimum weighing value, the
" " indication flashes.
When the amount of sample reaches the minimum weighing value or more,
the " " indicator will turn off
The minimum weighing value can be changed from the function setting. The
factory setting is 0 g.
If the set value is 0 g, no warning will be displayed even if the minimum weighing
warning function is ON. Also, a value greater than weighing capacity cannot be set
as minimum weighing value.
There are two kinds of warning display as follows
"Excluding near zero"
"Including near zero"
Near zero is within ± 10 digits of 0 g.
Setting procedure
1.
Hold down the SAMPLE key to display the function setting .
2.
Press the SAMPLE key several times to display .
3.
Press the PRINT key.
4.
will be displayed. Press the RE-ZERO key to change the
display from to (excluding near zero)
or
(including near zero).
5.
To change the setting of the minimum weighing value, proceed to 6.
If the minimum weighing value will not be displayed, press the PRINT
key.
6. Press the SAMPLE key to display .
7. Press the PRINT key.
8. Set the minimum weighing value. The minimum weighing value can
be changed by the following key operation.
RE-ZERO(+) key
Change the value of the blinking digit.
MODE
(
-
)
key
Change the value of the blinking digit.
SAMPLE key
Move the blinking digit.
PRINT key
Store the set value and proceed to the next item.
CAL key
Advance to the next item without storing the setting value.
9. Press the PRINT key to return to the weighing display.
10. Press the CAL key to return to the weighing display.