20
About the operation at when power is turned on
The balance will decide the reference zero point when the power is turned on (AC adapter is connected).
Depending on the load condition at that time, it will automatically judge whether to perform zero or tare
operation. The condition for determining which is used is "power on zero range", and when power on zero
range is exceeded, the tare subtraction operation is performed.
About re-zero operation
By pressing the RE-ZERO key, the display can be changed to zero.
Re-zero with the RE-ZERO key will automatically determine whether zero or tare operation is
performed.
The condition for determining which is used is "zero range", and when zero range is exceeded, the tare
subtraction operation is performed.
About measurement range
For the balance, the range that can be weighed is determined by model.
The total amount (net tare quantity) up to the maximum display of each model is displayed, and
when the maximum display is exceeded, E is displayed to indicate that the weighing range is
exceeded.
When in excess in negative, -E is displayed.
Model
Power on zero range
Zero range
-E display range
GX-203A
GF-203A
Approx.
±
100g
Approx.
±
4g
Approx.-100g or less
GX-303A
GF-303A
Approx.
±
100g
Approx.
±
6g
Approx.-100g or less
GX-403A
GF-403A
Approx.
±
100g
Approx.
±
8g
Approx.-100g or less
GX-603A
GF-603A
Approx.
±
100g
Approx.
±
12g
Approx.-100g or less
GX-1003A GF-1003A
Approx.
±
100g
Approx.
±
20g
Approx.-100g or less
GX-1603A GF-1603A
Approx.
±
100g
Approx.
±
32g
Approx.-100g or less
GX-2002A GF-2002A
Approx.
±
1kg
Approx.
±
40g
Approx.-1kg or less
GX-3002A
GF-3002A
Approx.
±
1kg
Approx.
±
60g
Approx.-1kg or less
GX-4002A GF-4002A
Approx.
±
1kg
Approx.
±
80g
Approx.-1kg or less
GX-6002A
GF-6002A
Approx.
±
1kg
Approx.
±
120g
Approx.-1kg or less
GX-10002A GF-10002A
Approx.
±
1kg
Approx.
±
200g
Approx.-1kg or less
GX-6001A
GF-6001A
Approx.
±
1kg
Approx.
±
120g
Approx.-1kg or less
GX-10001A GF-10001A
Approx.
±
1kg
Approx.
±
200g
Approx.-1kg or less