48
10-3
Description Of The Class "Environment, Display"
Condition (
Cond
)
This parameter is for sensitive response to the fluctuation of a mass value. Used for
powder target mass, weighing a very light sample or when quick response weighing is
required. After setting, the balance displays FAST .
This parameter is for stable weighing with slow response. Used to prevent a mass
value from drifting due to vibration or drafts. After setting, the balance displays
SLOW .
Stability band width (
5t-b
)
This item controls the width to regard a mass value as a stable value. When the fluctuation per second
is less than the parameter, the balance displays the stabilization indicator and outputs or stores the data
by function setting
(
dout, dataA
etc.
)
The parameter influences the "Auto print mode". Also, the minimum
display being displayed is 1 digit.
ex. If 0.01 mg display is selected by pressing the SAMPLE key on the GX-303A, 0.01 mg is 1 digit.
This parameter is used for sensitive response of the stabilization indicator. Used for
exact weighing.
This parameter ignores slight fluctuations of a mass value. Used to prevent a mass
value from drifting due to vibration or drafts
Hold function (
Hold
) (Animal weighing mode)
This function is used to weigh a moving object such as an animal. When the weighing data is over the
weighing range from zero and the display fluctuation is within the stabilization range for a fixed period of
averaging time, the processing indicator illuminates and the balance displays the average weight of the
animal. When the animal is removed from the weighing pan, the display returns to zero automatically. This
function is available only when the hold function parameter is set to "
1
" (the animal mode indicator ANIMAL
illuminates) and any weighing unit other than the counting mode is selected. The stabilization range and
averaging time are set in "Condition (
Cond
)" and "Stability band width (
5t-b
)".
Weighing range
Averaging time
Stabilation range
0.001g model
0.200g or more
Cond 0
2
sec.
(Efficiency priority)
4
sec.
8
sec.(Exact priority)
5t-b 0
0.01g model
2.00g or more
Cond 1
5t-b 1
0.1g model 20.0g or more
Cond 2
5t-b 2
※
Animal container kit (GXA-12) can be installed except GX-203A.
Zero tracking (
trc
)
This function tracks zero point drift caused by changes in the environment and stabilizes the zero
point. When the weighing data is only a few digits, turn the function off for accurate weighing.
trc
0
The tracking function is not used. Used for weighing a very light sample.
trc
1
The normal tracking function is used.(
±
1 digit / 1 second)
trc
2
The strong tracking function is used. (
±
1 digit / 0.5 second)
trc
3
The very strong tracking function is used. (
±
2 digit / 0.2 second)
〃
Lesser 6.25%
12.5%
Greater 16.7%