Error Codes
Display
Error code
Description
Data abnormality of the internal sensor of the balance
□
If this error occurs during weighing and you return to
the normal weighing state after a certain period of
time, there is a possibility that the data temporarily
became abnormal due to static electric noise if you
are measuring an easily chargeable item.
※
If the sample to be measured is charged, it is
recommended that you use a AD- 1683 static
eliminator to discharge the sample before
measurement in order to more accurately weigh
the sample.
□
If this error continues to be displayed and can not be
resolved, the internal sensor or the circuit may be
damaged. Please contact your dealer.
EC,E11
Stability error
The balance cannot stabilize due to an environmental
problem. Check around the pan. Prevent vibration, drafts,
temperature changes, static electricity and magnetic fields,
from influencing the balance. Refer to "2-3 During Use".
To return to the weighing mode, press the CAL key.
Out of the setting range
The data to be stored is out of the setting range.
Malfunction of the internal memory element of the
balance
If this error continues to be displayed, repair is necessary.
Please contact your dealer.
EC,E16
Internal mass error
Applying the internal mass does not yield a change in the
mass value as specified.
Confirm that there is nothing on the pan and perform the
weighing operation from the beginning again.
EC,E17
Internal mass error
The internal mass application mechanism does not
function properly. Perform the weighing operation from
the beginning again.
Abnormality in the internal memory data of the
balance
If this error continues to be displayed, repair is necessary.
Please contact your dealer.
Abnormality in the internal memory data of the
balance
If this error continues to be displayed, repair is necessary.
Please contact your dealer.
EC,E20
Calibration weight error
The calibration weight is too heavy. Confirm the calibration
mass value. Press the CAL key to return to the weighing