24
Notes
If improper performance is found in the self check, the balance displays
CH no
. Contact the
local A&D dealer for repair.
If the automatic response adjustment fails, the balance displays
CH ng
. Check the ambient
conditions such as breeze and vibration, also check the weigning pan. Then, perform the
adjustment again. To return to the weighing mode, press the CAL key.
If the automatic response adjustment is awkward, try to refine it using the manual response
adjustment.
6-2 Manual Response Adjustment
This function manually updates the response adjustment.
Operation
1 Press and hold the MODE key until RESPONSE is
displayed, then release the key.
And then, press the MODE key again quickly.
2 Press the MODE key to select a rate of the response
adjustment. Either FAST, MID. or SLOW can be
selected.
3 After a few seconds of inactivity the balance displays
e
end
. Then, it returns to the weighing mode and
displays the updated response indicator. The
response indicator remains displayed for a while.
Note
The response adjustment can be changed at “Condition
(
Cond
)” of “Environment, Display (
ba5fnc
)” in the
function table. For details, refer to “9. FUNCTION
TABLE”.