4-2 Changing the Units
The units or modes can be selected and stored in the function table. The sequence of displaying the
units or modes can be arranged so as to fit the frequency of use in the function table.
Select a unit or mode and arrange the sequence of display as follows:
1 Press and hold the SAMPLE key until
ba5fnc
of the
function table is displayed, then release the key.
2 Press the SAMPLE key several times to display
Unit
.
3 Press the PRINT key to enter the unit selection mode.
4 Specify a unit or mode in the order to be displayed
using the following keys.
SAMPLE key
To sequentially display the units.
RE-ZERO key
To specify a unit or mode. The
stabilization indicator appears when
the displayed unit or mode is
specified.
5 Press the PRINT key to store the units or modes.
The balance displays
end
and then displays the
next menu item of the function table.
6 Press the CAL key to exit the function table. Then the
balance returns to the weighing mode with the
selected unit.