24
10
-
2
Setting of the Decimal Point Position
The AD-8922A can set the display digit and the BCD output digit by setting
dpp
of the function.
When setting the decimal point position, the BCD output digit does not change if changing the minimum
display digit by pressing the SAMPLE key.
Example 1
)
When not setting the decimal point position
(
dpp -
)
[Factory setting]
(Key operation)
Balance display
AD-8922A display
BCD output
00012345
00123456
* When changing the minimum display digit by pressing SAMPLE key, the BCD shifts the output left and
adds the last digit.
Example 2
)
When setting the decimal point at the third digit position.
(
dpp 3
)
(Key operation)
Balance display
AD-8922A display
BCD output
00123450
00123456
* When changing the minimum display digit by pressing SAMPLE key, the BCD output does not change
the number of digits.