AstroPure 2000 LCD 
 

To access the user menu (to unlock the display), the user can press the ‘AstroPur’ logo or, 
alternatively, press the lock symbol for 2 seconds.

The user menu password is “116”.

Main User Menu

Press the AstroPure logo for 3 seconds.

The language settings can be modified and system information can be viewed 
(software version, hardware ID).

After 60 seconds of inactivity, the main screen will automatically ‘lock’, viz. no 
changes can be made to any settings, e.g., main standby switch on/off, airflow set 
point, lamp on/off, etc.

To modify any setting, the screen needs to be unlocked again.

After 60 seconds of inactivity, the main screen will automatically ‘lock’, viz. no 
changes can be made to any settings, e.g., main standby switch on/off, airflow set 
point, lamp on/off, etc.

To modify any setting, the screen needs to be unlocked again.

Operation (LCD / PLC Control)

• 

Plug in the power plug

• 

Set the ON/OFF switch at the back side of the unit to ON.

 

 

Set the desired volume flow.

• 

Press the standby button to turn on the fan 

 

•  To unlock the screen, enter the password “116”

• 

Volume flow adjustment:

• 

Method (1)

• 

Press the flow rate bar to set the initial value. The exact desired speed of the fan 

can be set here. 

• 

Press the plus and minus button. The respective speed increment is set in the 

configuration menu.

• 

Method (2)

• 

Press an area on the flow rate bar. With this option, the speed of the fan is set to 

one of the 4 parameter values at a time (25% steps).

• 

Information on display:

Service: filters must be changed by qualified personnel

Configuration menu, language and system information

Press the AAF-logo field for 5 seconds to enter the configuration menus.

Enter the password “-1101“.

The controller‘s parameters can be changed here.

Resetting of operating hours

In the configuration menu, these values can be reset by pressing the corresponding button for 

5 seconds.

Appropriate flow rate

To achieve best air cleaning results, the air cleaners flow rate must be adjusted to the individual 
situation of the designated room.

Following calculation steps provide a basic guide line:

•  The volume of the designated room in cubic meters (m³) is calculated by multiplying the room’s 
length * width * height in meters: V = L * W * H

•  The desired number of room air exchanges per hour (ACH) has to be selected (units: 1 /h). This 
number is sometimes also referred to as air exchange rate.

When this value is not preset, the following ACH values represent the currently applied 
recommendations:

•  Good ventilation:  

 

ACH = 4 1/h

•  Excellent ventilation:  

 

ACH = 6 1/h

•  Purging of rooms:  

 

ACH = 10 1/h

16

 AstroPure 2000 Main menu

-> after 5 s ->

Summary of Contents for AstroPure 2000

Page 1: ...AstroPure 2000 Air Cleaner IOM AstroPure Installation and Operation Manual...

Page 2: ...transport and assembly 13 Commissioning Operation Cleaning and Disposal 19 Emergency 20 Drawings and diagrams 21 Spare parts 21 Certificates and Protocols 22 Appendix Potentiometer Control Manufacture...

Page 3: ...be trained and instructed for example according to VDI 6022 4 The AstroPure air purifier is shipped with an installation operation and maintenance manual AAF HEPA filter test report For maintenance i...

Page 4: ...switch in OFF Set the speed controller to the OFF position Plug in the power plug ON OFF switch ON Check serial number for fan type on the Type Plate on the back of the uni Serial No E11 2020 12345 E...

Page 5: ...teristics example below Image 1 AstroPure type label 3 Quick Start Guide LCD PLC Control Quick start guide Observe the safety instructions for safe operation Ensure a safe stand on smooth level stable...

Page 6: ...Device weight kg 150 Min ambient temperature C 0 Max ambient temperature C 40 Allowed rel humidity level 0 90 depending on execution Connection details and power consumption Connection details 1PH 23...

Page 7: ...ackaging Storage temperature range 30 C to 40 C Air humidity rH 90 no condensable humidity Unpacking Remove the securing straps stretch wrap and card board packing material from the AstroPure 2000 Air...

Page 8: ...vice is properly closed Make sure that the air inlet and outlet openings are freely accessible and not obstructed or occupied Fix the locking devices of the castors General In case of any abnormal noi...

Page 9: ...be set here Press the plus and minus button The respective speed increment is set in the configuration menu Method 2 Press an area on the flow rate bar With this option the speed of the fan is set to...

Page 10: ...ied below or earlier if necessar 18 Parameter Symbol Unit Example 1 Example 2 Example 3 Example 4 Length L m 10 10 10 10 Width W m 10 10 10 10 Height H m 2 2 5 2 5 2 5 Volume V m3 200 250 250 250 Air...

Page 11: ...enwaldstrasse 4 D 64646 Heppenheim Germany Telefon 49 0 6252 69977 0 Email Sales DACH aafeurope com A risk assessment and compliance evaluation procedure has bend conducted referenced on the following...

Page 12: ...LLED PRODUCT OWNER FM AAF DRAWING NO mm SHEET UL DO NOT SCALE DRAWING REF Material Finishing MATER Roughnes 1 1 2 2 3 3 4 4 5 5 6 6 7 7 8 8 A A B B C C D D E E F F AP2000 LCD Guillaume Fogeron 19 03 2...

Page 13: ...10 663 51 09 E mail greece aafeurope com Italy AAF S r l Via Friuli 28 30 21047 Saronno VA Phone 39 02 9624096 Fax 39 02 9606409 E mail sales italy aafeurope com Dinair Norway Dinair AS Prof Birkeland...

Reviews: