55
GlucoMen LX PLUS - 42338 - 09/10
UK
System Hardware Error.
Call A. Menarini Diagnostics for
assistance:
(UK) 0800 243667
(Republic of Ireland) 1800 709903
(Northern Ireland) 0800 7837286
Operating Temperature Error.
Meter is outside the required testing
temperature range of 5 to 45 °C (41 to 113
°F). Move meter and sensors to a warmer
or cooler area and wait a few minutes.
Used Sensor Error: used or damaged
sensor.
Retest with a new sensor.
Blood Sample Error.
Incorrect application of blood sample or
control solution onto the sensor, or the
sensor may be damaged. Review your
sampling technique. The blood sample
quantity may be incorrect or hematocrit
may be out of the allowed range for
glucose measurement.
11. DISPLAYS, MEANINGS, ACTIONS
11. DISPLAYS, MEANINGS, ACTIONS
Display
What it Means
What to Do