68
3M
™
Steri‑Vac
™
Sterilizer / Aerator GSX Series
– Operator’s Manual
0
0
40
80
120
160
200
20
40
60
80
100
0
20
40
60
80
100
minutes
%RH
°C mbar/10
Figure 81
Example Cycle Report – Graph Option
Stage
Preheat
Air removal
Chamber test
Conditioning
EO injectin
EO exposure
EO removal
Flushing
Aeration
Tm (C)
21.2
54.0
54.6
54.7
55.0
56.1
55.0
55.0
55.0
Tmd (C)
21.1
53.5
54.0
54.2
54.4
54.4
54.4
54.4
54.5
Pc (kPa)
97.06
77.49
15.95
16.22
15.38
61.58
60.02
16.79
79.70
HS (C)
31.0
54.9
94.7
94.7
94.9
61.7
54.9
55.0
54.9
RHc (%)
5.2
59.1
RHm (%)
5.1
56.4
Pm (kPa)
97.14
77.60
15.98
16.25
15.41
61.64
60.12
16.78
19.85
Tul (C)
21.1
54.3
54.9
55.1
55.3
558.3
55.2
55.2
55.1
Tld (C)
21.0
54.2
54.7
54.9
55.1
56.1
55.0
55.0
54.9
Tc (C)
21.2
54.0
54.6
54.7
55.0
56.0
55.0
55.0
55.0
Start time
03:10:05
03:39:19
03:49:19
03:55:22
05:03:28
05:05:00
06:05:02
06:18:54
08:40:43
Figure 82
Example Cycle Report – Table Option