11.7. Running Custom Cycles on the 3M™ Steri‑Vac™ Sterilizer/Aerator GSX Series
DANGER: To reduce the risks associated with exposure to ethylene oxide:
Short term exposure limits (STEL) or long term exposure limits (LTEL) or immediately dangerous to life or health (IDLH) limit could be exceeded during NORMAL use Operators and staff
must use protective measures (eg engineering controls, work practices, or personal protective equipment (PPE)) in accordance with United States 29 CFR 1910134 and 29 CFR
19101047 under NORMAL use conditions until the user’s facility management completes an exposure assessment on each custom cycle that validates facility protective measures meet
national, state, and local requirements for short term exposure limits (STELs) and long term exposure limits (LTELs)
Assure Operator selects the correct cycle for the intended application
CAUTION: To reduce risks associated with exposure to potentially non‑sterile devices or inadequate sterilization procedures:
The user must validate the sterile efficacy of each custom cycle intended to sterilize products labeled ‘Sterile’
Assure Operator selects the correct cycle for the intended application
1�
A Supervisor must access
Menu>Site Setup>Setup>Cycle
to enable newly imported cycles on the Operator Cycle Start Screen Newly imported cycles are not automatically added to the
Operator Cycle Start Screen view (Figure 57)
Note
:
• Cycles created for the Operator will be selectable in Operator view
• Cycles created for the Supervisor will be selectable in Supervisor view
• Once selections are made, press
Apply Changes
Press
Return
to go back to the main screen
Note:
• This feature allows custom cycles to be stored and accessed in different locations on the GSX Series sterilizer to help assure Operator selects and starts the correct cycle for the
intended application
• Minimum aeration times can not be changed in this view for custom cycles Minimum aeration times must be changed in the 3M™ Cycle Programmer and then imported back to the
GSX Series sterilizer
Sterilizer
22.2
˚
C
04/20/2015
14:12:30
98.0
kPa
Select cycle temperature:
6
5
4
3
2
1
7
Return
Apply changes
Cycle access level
Touch check box to select cycles listed on start screen.
Touch row to change values.
Operator
Cycle
GSX 38C
GSX 55C
36.0 hrs
12.0 hrs
Minimum aeration
Scope 395
Stent 126
Setup
Preferences Cycle Setup Date-Time Network Site
Figure 57
Imported Cycles Stored in
Menu>Setup >Site Setup>Cycle