84
3M
™
Steri‑Vac
™
Sterilizer / Aerator GSX Series
– Operator’s Manual
Once the automatic error recovery cycle is complete, a larger message will appear on the display screen as illustrated in Figures 98 and 99
E702 - User abort
Details...
Clear
04/20/2015 04:22:10
1. Error Number
2. Error Description
3. Error
Clear Button
4. Error Additional Details
Toggle Button
5. Error Date and Time
Figure 98
Error Message –Details Hidden
E702 - User abort
04/20/2015 04:22:10
Error level: L4
Stage : EO exposure
Info: User aborted cycle
Hide
Clear
6. Error Level
7. Error Stage
8. Error Information
Figure 99
Error Message – Details Revealed
Item
Number
Error Name
Error Item Description
1
Error Number
Alpha‑numeric designator of the error
2
Error Description
Text description of the error This description will be helpful information for 3M Health Care service personnel or authorized
3M service personnel
3
Error Clear Button
Press this button to acknowledge and clear the error Some errors must be cleared by 3M Health Care service personnel or
authorized 3M service personnel with a Service PIN See Table 8 – Descriptions for Error Levels and Corrective Actions for
more details
4
Error Additional Details Toggle Button
Press toggle button to reveal the Error Level, sterilization cycle stage, and additional information regarding the error
5
Error Date and Time
The date and time the error occurred
6
Error Level
Error is divided into levels Each level has distinct automatic safe recovery actions and Operator corrective actions See Table 8 ‑
Descriptions for Error Levels and Corrective Actions for more details
7
Error Stage
Stage of the sterilization cycle or aeration stage when the error occurred
8
Error Information
Additional information regarding the error which will be helpful 3M Health Care service personnel or authorized 3M
service personnel
Table 9. Explanation of Error Messages