19
EO usage report
AA 121212 04/20/2015 12:30:55
Software version 1.0.1
----------------------------------
Sterilizer ID = Machine 2
Start date = 04/19/2015
End date = 05/01/15
Total cycles = 10
Total EO weight (kg/lb) = 1.00/2.20*
*The documented Total EO weight is based on the
NOMINAL fill weight of 100% Ethylene Oxide. The ACTUAL
fill weight may be different. ACTUAL fill weight met all 3M
specifications at the time of product release.
----------------------------------
Figure 14.
Printed EO Usage Report
10.3.3. Site Setup Report
View a Site Setup Report for the 3M™ Steri-Vac™ Sterilizer/Aerator GS Series by viewing, printing, or exporting electronically to a USB drive via
Menu>Reports>Site Setup Report
. The Site
Setup Report contains the site settings established in the Site Setup menu. Figure 15 is the display screen view of the Site Setup Report.
Sterilizer
22.2
˚
C
04/20/2015
12:30:30
98.0
kPa
Select cycle temperature:
6
5
4
3
2
1
7
55 °C
55 °C
Site setup report
Export
Export
Cancel
Cancel
Site setup report
AA 121212 04/20/2015 12:32:52
Software version 1.0.1
_________________________
Preferences
Audible tone on cycle failure = Yes
Audible tone on cycle complete = Yes
Print on cycle complete = No
Cycle report preference = Graph
Sensors to plot = Control
Screen timeout = 10 minutes
Idle mode chamber temperature = 35 ˚C
Pressure unit = kPa
User count = 38
Abated aeration time = 0.0 hours
Setup
Abated aeration time
t
Figure 15.
Display Screen View of the Site
Setup Report