9
7. Specifications
7.1. 3M™ Steri-Vac™ Sterilizer/Aerator GS Series Structural Specifications
GS5
GS8
Shipping Weight
163 kg (359 lbs.) single door
169 kg (373 lbs.) double door
355 kg (782 lbs.) single door
362 kg (799 lbs.) double door
Operational Weight
127 kg (281 lbs.) single door
132 kg (290 lbs.) double door
261 kg (576 lbs.) single door
269 kg (593 lbs.) double door
Exterior Dimensions
H 70.9 cm x W 76.2 cm x D 95.0 cm
H 27.9 in. x W 30.0 in. x D 37.4 in.
H 179.8 cm x W 94.0 cm x D 109.0 cm
H 70.8 in. x W 37.0 in. x D 42.9 in.
Chamber Internal Volume
136 L (4.8 cubic feet)
224 L (7.9 cubic feet)
Chamber Internal Dimensions
H 38.0 cm x W 43.0 cm x D 83.0 cm
H 15.0 in. x W 17.0 in. x D 32.5 in.
H 46.0 cm x W 51.0 cm x D 97.0 cm
H 18.0 in. x W 20.0 in. x D 38.0 in.
7.2. Sound Levels
The average decibel level of an active GS5 or GS8 sterilizer is <60 dBA. For more information on sound levels, in the US contact 3M Health Care Helpline at 1-800-228-3957. Outside the US
contact your local 3M office or to locate your local office go to www.3M.com
7.3. Power Specifications
WARNING: To reduce the risk of shock due to hazardous voltage,
the customer must provide a properly grounded outlet (an earth ground) for installation as described in the installation requirements section of this manual. Do not use a detachable cord rated
less than 15A.
CAUTION: To reduce patient risks associated with exposure to potentially non-sterile devices or inadequate sterilization procedures,
do not place any device emitting strong electronic magnetic fields (EMFs) near the sterilizer.
Electrical Power
Operating Condition
Units
Voltage Range
200 - 240
VAC
Frequency
50/60
Hertz
Phase
Single
Not Applicable
GS5 Current
10
Amps
GS8 Current
12
Amps