10.3. Reports
The 3M™ Steri-Vac™ Sterilizer/Aerator GS Series has a variety of reporting options as shown in Figure 10.
Sterilizer
22.2
˚
C
98.0
kPa
Select cycle temperature:
04/20/2015
05:19:53
6
5
4
3
2
1
7
Reports
Cycles
Setup
Status
Service
Cycle reports
EO usage report
Site setup report
Printer form feed
55 °C
Figure 10.
Reports Menu
10.3.1. Cycle Reports
Figure 11 outlines the options for 3M™ Steri-Vac™ Sterilizer/Aerator GS Series Cycle Reports.
Sterilizer
22.2
˚
C
04/20/2015
12:29:13
98.0
kPa
Print or export report
Print selected cycle
Export selected cycle
Export all cycles
Pressure
Relative humidity
Pressure
Relative humidity
Monitor sensors
Control sensors
Temperature
Report format
OK
OK
Return
Return
Enter Pin
Enter Pin
Temperature
Graph
Cycle 33 - 4/20/2015 - 55 ˚C
Cycle 32 - 4/19/2015 - 38 ˚C
Cycle 31 - 4/18/2015 - 55 ˚C
Cycle 30 - 4/17/2015 - 55 ˚C
Cycle 29 - 4/16/2015 - 55 ˚C Aeration
Cycle 28 - 4/15/2015 - 55 ˚C
Cycle 27 - 4/12/2015 - 38 ˚C
Cycle 26 - 4/11/2015 - 55 ˚C
Cycle 25 - 4/10/2015 - 55 ˚C
Figure 11.
Cycle Report Setup
Reports from the last 100 cycles can be printed or exported electronically to a USB drive. When the 100 cycle limit is reached, the oldest cycles are replaced with the most recently run cycles.
Select the desired cycle(s) from the list of the last 100 cycles (sorted with the most recently run cycles at the top) and select the desired function from the following options: print selected cycle,
export selected cycles, or export all cycles. Only one cycle may be selected for printing. Multiple cycles can be selected for export to a USB drive.
To export electronic reports, insert a USB drive (data storage device) into one of two USB ports on the GS Series sterilizer. Recommended USB drives include those with FAT32 formatting. USB
drives with pre-loaded software (e.g. SanDisk’s Cruzer®) are not recommended. Connect only USB drives for the export of data to the USB ports. Do not connect external USB devices that supply
power to the USB ports.
Select the cycle(s) for which you wish to export reports and press the
OK
button. The selected reports are stored in the USB directory selected by the user. The sterilizer will ask for confirmation
of the Folder designation before export. The electronic reports are generated in color and are sized as 20.3 cm (8 in.) x 27.9 cm (11 in.) images and contain the same information as the strip
chart reports.
The GS Series sterilizer has control sensors for temperature, %RH, and pressure that provide information to the control embedded software. In addition, the sterilizer has a duplicate set of
monitoring sensors that provide independent data and performance monitoring during critical sterilization stages for temperature, %RH, and pressure. The sterilizer defaults to report only on the
control sensors. Monitoring sensors can be selected for reporting in the Cycle Report options screen in Figure 11.