background image

 

 

 

 

1.

 

Ensure the SoundPro is turned On and is either stopped or paused. 

2.

 

Attach the calibrator and cal adapter to the SoundPro.  Set the calibrator to 1 KHz and 114 dB (if it is a selectable). 

 
 
 

 
 
 
 
 
 
 
 
 

 
 

3.

 

From the Start screen, press the 

Cal

 softkey and the Cal screen will open.  Press         when Calibrate is highlighted.  The Pre-Cal 

screen will open. 

 
 
 
 
 
 

 

4.

 

Allow the measurement to stabilize and press             

Arrows

 to set 114.0 dB level (see A).  Then, press         key.  When set, the 

Calibration history screen appears (see B).   

 

 

 

 
 
 
 
 
 
 

5.    Press 

         

key to exit the Cal screen. 

  

6.    For a 

Post Calibration

, ensure the SoundPro is in “Pause” mode.  (Essentially, run a measurement and press the run/pause key.)  

The Pause indicator will appear at the top of the display.)  Then repeat the Calibration steps above and press Enter to store the Post 
Calibration information.  A sample screen is displayed below with post calibration history. 

 
 
 
 
 
 
 

 

 

 

Position the SoundPro in the field.  Press the run key to start measuring.  To pause a study (for example:  lunch breaks), press the run/pause key. To 

 

start measuring again, press the run/pause key.  Continue this process until your measurements are complete by press the Stop key.  This will end 

 

the session.   

1.   From the 

Start screen

, select either 

View Session

 or select 

View Current Study

 by press the              arrows and         key to select.   

 

2.   Press          key to begin measuring.  Press          key to Pause (optional).  At any time, press          to end the session/study.   

 

 

 

 

QC10/QC20 

Select 1kHz and 114 dB  

Microphone 

Place calibrator 
over microphone 

 

A. Calibrate 

Press corresponding 
softkey to 114.0 dB 
level. 

 

Cal softkey 

  Select softkey to open 

B.  Pre-Cal 

Displays pre-calibration 
information 

RunTime 

Run icon 

Lookup softkey:  
Select to change 
measurement 
parameter (Lpk, 
Lmx, Lmn, 
Leq,Las) 

Measurement value (Average SPL) 

Summary of Contents for SoundPro SE Series

Page 1: ...3M Personal Safety Division 3MTM SoundPro SE DL SeriesSound Level Meters SoundProQuick Start Guide ...

Page 2: ... settings to fit the specifications of your sound study To open the Setup menu from the start screen press the arrow keys and press key when selected The setup menu will appear To change the parameters navigate to the appropriate menus and make changes as needed See the user manual for specific details and term explanations for more information 1 Display 8 Softkeys 12 On Off and Esc Escape key 2 B...

Page 3: ...re the Post Calibration information A sample screen is displayed below with post calibration history Position the SoundPro in the field Press the run key to start measuring To pause a study for example lunch breaks press the run pause key To start measuring again press the run pause key Continue this process until your measurements are complete by press the Stop key This will end the session 1 Fro...

Page 4: ... measured value is above the minimum value for the selected measurement range Community Noise screen This screen shows the results of measurements commonly associated with community noise studies Dosimetry screen The Dosimetry screen provides several types of measurements of the accumulated personal exposure to noise Broadband amplitude Amplitude in the band Selected band Broadband Bar Chart view ...

Page 5: ...trument communication A Click on icon see A to access the Instrument download panel 4 Selecting and viewing data A The Data Finder page see A is used to view select and preview reports Note Click Browse to view all data files B Select the data by clicking on either the session study Note you may need to expand to see the study data C To view in Graphs and Tables click on or buttons B A Data Finder...

Page 6: ...ality and select toxic combustible gases The 3M Detection brand of instrumentation is used by safety and industrial hygiene professionals to help comply with applicable occupational standards and regulations About 3M Personal Safety 3M offers a comprehensive diverse portfolio of Personal Safety solutions providing respiratory protection hearing protection fall protection reflective materials for h...

Reviews: