3M™ PROTÉGÉ SG
PORTABLE, REUSABLE, SINGLE-GAS MONITOR
CONFIGURATION AND SETUP
087-0065, Rev D/February 2019
4-6
4.5. Monitor Setup on PGD Application Software
You can set up the User Options Menu via the PGD Application software.
lists the functions that can be enabled or disabled through the software.
Perform the following steps to set the monitor parameters:
CO (Hydrogen
Resistant)
0 - 2000ppm
35ppm
200pp
m
35ppm
200pp
m
50ppm CO
HCN
0 - 100ppm
10ppm
4.7ppm
4 ppm
10ppm
10ppm HCN
O
3
0 to 3 ppm
0.1 ppm
0.1 ppm
0.1 ppm
0.5 ppm 2 ppm Cl
2
NO
0 to 350 ppm
25 ppm
N/A
10 ppm
25 ppm
25 ppm NO
NO
2
0 to 100 ppm
3 ppm
5 ppm
2.5 ppm 5 ppm
5 ppm NO
2
PH
3
0 to 20 ppm
0.3 ppm 1 ppm
0.3 ppm 1 ppm
1 ppm PH
3
SO
2
0 to 150 ppm
2 ppm
5 ppm
2 ppm
5 ppm
10 ppm SO
2
NH
3
0 to 300 ppm
25 ppm
35 ppm
25 ppm
50 ppm
50 ppm NH
3
Cl
2
0 to 50 ppm
0.5 ppm 1 ppm
0.5 ppm 1 ppm
5 ppm Cl
2
ClO
2
0 to 5ppm
0.1 ppm
0.3 ppm 0.1 ppm
0.3 ppm 2 ppm Cl
2
Table 4-3: Firmware Default Settings
Options
Range
Factory Default
Backlight Duration
2-255 seconds
30 seconds
Calibration Period
1 to 365 days
180 days
Bump Test Period
1 to 365 days
180 days
Compliance LED
1 ~ 255 seconds
60 seconds/ON
Silent Alarm
1 ~60 seconds
60 seconds/ON
On Timer
3 ~9 seconds
3 seconds
Off Timer
3 ~9 seconds
5 seconds
Button Hold Timer (Test)
3 to 9 seconds
3 seconds
Confidence Beep
1 ~ 255 seconds
60 seconds/OFF
Auto Zero Calibration on Start-up
ON/OFF
OFF
Enable Latching Alarms
ON/OFF
OFF
Table 4-2: Factory Default Alarm Setpoints
Gas
Range
TWA
STEL
Low
High
Gas Type &
Value for CAL