General Maintenance
Calibrating the Operating Temperatures
Calibrate the operating temperatures of the Model 875 warming unit (1) after performing
a service procedure and (2) routinely every six months� Perform the calibration in a similar
environment temperature to where the Model 875 will be used� You will need a Bair Paws
temperature test kit Model 90055 to perform the calibration� Calibration instructions are
included with the test kit�
Replacing the Filter
Replace the filter every six months or sooner if necessary� To order a replacement filter,
contact 3M Patient Warming customer service using either of the telephone numbers listed
on page 22� Instructions for replacing the filter will be included with the replacement filter�
Cleaning the Warming Unit, Handheld Controller, Hose, and Accessories
NOTICE:
To avoid device damage:
•
Do not immerse any part of the Bair Paws warming unit in liquid�
•
Do not use a dripping wet cloth to clean the warming unit�
•
Do not use harsh solvents to clean the warming unit�
Method
1�
Disconnect the warming unit from the power source before cleaning�
2�
Wipe the cabinet, the temperature controller, and the outside of the hose with a damp,
soft cloth and a mild cleaning solution or antimicrobial spray�
3�
Dry with a separate soft cloth�