WARNING:

To reduce the risk associated with incorrect results

Instrument is to be used by operators familiar with the device, device functionality, and

Operator

s Manual.

Do not place the instrument in environment exposed to sunlight or strong

incandescent light.

Do not place instrument close to any device that emits a strong electro

magnetic field.

Do not use on a vibrating surface.

Activate and incubate 3M™ Attest™ Rapid Readout Biological Indicators 1295 within

1 hour of the completion of the sterilization cycle.

Do not incubate a 3M™ Attest™ Rapid Readout Biological Indicator 1295 if, after

processing and before BI activation, it is observed to have a broken media ampoule.

Retest the sterilizer with a new biological indicator.

Incubate the 3M™ Attest™ Rapid Readout Biological Indicator 1295 only in the

3M™ Attest™ Auto

reader 490H i.e., match the pink cap color of the 1295 BI with

the pink color

coded configuration sticker surrounding the 3M™ Attest™ Auto

reader

490H incubation wells.

Do not remove the 3M™ Attest™ Rapid Readout Biological Indicator from the

incubation well until the (+) or (

) symbol on the LCD display indicates the test

is complete.

Caution:

To reduce the risk associated with incorrect results

To avoid the possibility of the 3M™ Attest™ Rapid Readout Biological Indicator

1295 vial absorbing fluorescent residue from a chemical indicator or tape, place the

3M™ Attest™ Rapid Readout Biological Indicator vial so it does not come in direct

contact with chemical indicators or tape.

HEALTH AND SAFETY INFORMATION

Device Safety Compliance

The 3M™ Attest™ Auto

reader 490H complies with the following standards as

demonstrated by the CB Scheme Certificate and test report issued by Underwriters

Laboratories (UL):
• IEC 61010

1 (2010) 3rd Edition. Safety requirements for electrical equipment for

measurement, control, and laboratory use

 Part 1: General requirements

• IEC 61010

2

010: 2014 3rd Edition. Safety requirements for electrical equipment for

measurement, control, and laboratory use

 Part 2

010: Particular requirements for

laboratory equipment for the heating of materials

The 3M™ Attest™ Auto

reader 490H is listed as Laboratory equipment and carries the

UL mark with adjacent indicators

C

 and

US

 based on compliance to the standards UL

61010

1, CAN/CSA C22.2 No. 61010-1 and CAN/CSA C22.2 No. 61010-2-010.

The 3M™ Attest™ Auto

reader 490H complies with the CE mark related to the Low

Voltage Directive (LVD) 2006/95/EC as confirmed in the Declaration of Conformity.
The 3M™ Attest™ Auto

reader 490H complies with the RoHS Directive, Directive

2011/65/EU of the European Parliament and of the Council of 8 June 2011 on

the restriction of the use of certain hazardous substances in electrical and

electronic equipment.
The 3M™ Attest™ Auto

reader 490H complies with the WEEE Directive, Directive

2012/19/EU of the European Parliament and of the Council of 04 July 2012 on waste

electrical and electronic equipment (WEEE).

EMC Compliance

The 3M™ Attest™ Auto

reader 490H complies with the following EMC standards as

confirmed in the Certificate of Compliance generated by 3M:
• IEC 61326

1:2012/EN 61326:2013 Electrical equipment for measurement, control and

laboratory use

 EMC requirements

 Part 1: General requirements

• EMC requirements of the CE mark EMC Directive 2004/108/EC.

The 3M™ Attest™ Auto

reader 490H complies with the Australian and New Zealand

electrical safety and electromagnetic compatibility requirements as confirmed in the

Supplier

s Declaration of Conformity that is linked to the Australian/New Zealand RCM

(Regulatory Compliance Mark).

Note:

This equipment has been tested and found to comply with the limits for a Class A

digital device pursuant to Part 15, Subpart B, of the FCC Rules. These limits are designed

to provide a reasonable protection against harmful interference when the equipment is

operated in a commercial environment. This equipment generates, and can radiate radio

frequency energy and, if not installed and used in accordance with the instruction manual,

may cause harmful interference to radio communications. Operation of this equipment

in a residential area is likely to cause harmful interference in which case the user will be

required to correct the interference at their own expense.
This Class A digital device meets all requirements of the Canadian Interference

Causing

Equipment Regulations.

EXPLANATION OF PRODUCT AND PACKAGING

LABEL SYMBOLS

Attention

 Refer to the instructions for use

Waste Electrical and Electronic Equipment (WEEE) and EU Battery Directive.

This symbol indicates that both the device and the lithium ion battery

contained therein need to be disposed of properly.

UL Listed to US and Canadian Safety Standards

Mark of Conformity to European Directives

Direct current

Compliant to all applicable ACMA regulatory arrangements (RCM)

Manufacturer

Catalog number

Authorized representative for the European Community

Serial number

Li

REF

Summary of Contents for Attest Auto-reader 490H

Page 1: ...io 5Manual del operador 6Bedieningshandleiding 7Användarhandbok 8Brugsvejledning 9Brukerhåndbok 0Käyttöopas Manual do Utilizador Manual do Operador Εγχειρίδιο Χειριστή HInstrukcja obsługi FPříručka pro obsluhu SPriročnik za uporabo MИнструкция по эксплуатации Upute za korištenje dOperatör Kılavuzu I操作指南 jคู มือปฏิบัติงาน T작업자 매뉴얼 ...

Page 2: ......

Page 3: ...2 Product Description 2 Warranty 2 Safety and Precautions 2 Health and Safety Information 3 Explanation of Product and Packaging Label Symbols 3 Specifications 4 Instructions for Use 4 Features 7 Care and Cleaning 8 Troubleshooting Guide 8 Optional 3M Attest Auto reader 490H WebApp 9 Troubleshooting the 3M Attest Auto reader 490H WebApp 12 Repair and Replacement 12 ...

Page 4: ...fluorescence reading or a positive visual pH color change result indicate a sterilization process failure However due to the high sensitivity of the fluorescent result there is no advantage to continuing to incubate processed 3M Attest Rapid Readout Biological Indicators 1295 once the fluorescent result has been documented Other biological indicators BIs are not compatible with this device and can...

Page 5: ...H complies with the CE mark related to the Low Voltage Directive LVD 2006 95 EC as confirmed in the Declaration of Conformity The 3M Attest Auto reader 490H complies with the RoHS Directive Directive 2011 65 EU of the European Parliament and of the Council of 8 June 2011 on the restriction of the use of certain hazardous substances in electrical and electronic equipment The 3M Attest Auto reader 4...

Page 6: ...sed to sunlight or strong incandescent light Do not place instrument close to any device that emits a strong electro magnetic field Do not use on a vibrating surface Select Install Correct Power Plug The 3M Attest Auto reader 490H power supply comes with multiple blade configurations Install the blade configuration appropriate for your country into the power supply and plug into appropriate AC gro...

Page 7: ...etting Setting the 3M Attest Auto reader 490H Clock The 3M Attest Auto reader 490H date and time are displayed in this format YYYY MM DD HH MM with the time displayed using a 24 hour clock To change the year if necessary push the up or down buttons to set desired year then press to advance to the month adjust as necessary and press to advance to the day Use the button to move to the time and set t...

Page 8: ...ce the result is determined it is continuously displayed on the LCD panel until 10 seconds after the BI has been removed from the incubation well Note After the first minute of incubation if 3M Attest Rapid Readout Biological Indicator is inadvertently removed before the incubation period is complete an alarm will sound and a C2 Caution code will be displayed on the LCD panel under that well You m...

Page 9: ... positive 3M Attest Rapid Readout Biological Indicator results subsequently detected in other wells will result in the alarm sounding Removing the 3M Attest Rapid Readout Biological Indicator will reactivate the alarm for that well Pressing the Buzzer Alarm OFF button will not silence the alarm for error or caution code alarm warnings To disable the alarm see configuration section Stackable To sav...

Page 10: ...unit and plug it back in to clear error If error doesn t clear call for repair or replacement E3 Memory Error 3M Attest Auto reader 490H unable to save read data from memory Unplug the unit and plug it back in to clear error If error doesn t clear call for repair or replacement E7 Communications Error This error will occur if internal communications have failed Unplug the unit and plug it back in ...

Page 11: ...ddress URL Note To access the 3M Attest Auto reader 490H Web App within a secure network the PC or mobile device must have access to the secure network options include connecting via a network Ethernet cable signing in to the facility s wireless network or connecting via a virtual private network VPN A Serial number URL 1 Connect the supplied Ethernet cable from the 3M Attest Auto reader 490H to a...

Page 12: ...er 490H Web App Status History 3M Attest Auto reader 490H ATTEST 1295 ATTEST 1295 Auto reader 490H 1 2 3 4 5 24 6 7 8 9 10 SN 200299 2017 05 11 17 13 1 Open the Results Detail page After inserting a new 3M Attest Rapid Readout BI in the 3M Attest Auto reader 490H it is a best practice to open the 3M Attest Auto reader 490H Web App Results Detail page for that BI by clicking on the appropriate well...

Page 13: ...k on the desired radio buttons to change settings Language Selected setting indicates language in which text on 3M Attest Auto reader 490H Web App pages will appear on the PC mobile device screen Pushbutton Configuration Selecting Enable allows the user to configure the 3M Attest Auto reader 490H from the buttons on the unit itself Selecting Disable prevents the user from being able to configure t...

Page 14: ...cility virtual private network VPN on the mobile device Data entered by user on Results Detail page disappears User entered data is written to temporary web browser storage Browser may have cleared data Set web browser to accept allow or enable cookies As a best practice print or save Results Detail record as soon as BI result is final BI result displayed on 3M Attest Auto reader 490H LCD display ...

Page 15: ......

Page 16: ...t Paul MN 55144 1 800 228 3957 3M com Medical 3M Deutschland GmbH Health Care Business Carl Schurz Str 1 41453 Neuss Germany 2017 3M All rights reserved 3M and Attest are trademarks of 3M Used under license in Canada Issue Date 2017 03 34 8720 4946 4 ...

Reviews: