INSTRUCTIONS FOR USE

WELLS, BUTTONS, & DISPLAY INTERPRETATION

Please familiarize yourself with the face of the 3M™ Attest™ Auto

reader 490H:

Wells

1. Incubation/Reader Wells

 The 3M™ Attest™ Auto

reader 490H has ten

incubation wells.

Buttons

2. Information / Up Arrow Button

3. Temperature / Down Arrow Button

4. Buzzer Alarm OFF / Enter Button

Information Display

5. LCD display

 Display is backlit

6. LCD display

 Example of Error Message for well #6

7. LCD display

 Example of Remaining Incubation Time, in minutes, for well #10

8. LCD display

 Example of Negative (

 ) BI result for well #9

9. LCD display

 Example of Positive ( + ) BI result for well #8

10. LCD display

 Alarm Enabled Indicator

Color

Coded Configuration Indicator

11. Incubation well configuration sticker is color

coded pink

INSTRUCTIONS FOR USE

3M

ATTEST

AUTO

READER 490H CONFIGURATION

Note:

 Ideally, the 3M™ Attest™ Auto

reader 490H should be empty (i.e., have no

incubating BIs) when it is reconfigured.

Press the

and

buttons on the front of the 3M™ Attest™ Auto

reader 490H

simultaneously for 3 seconds to enter configuration mode. Once in configuration mode,

use the Up

 and Down

 buttons to set the 3M™ Attest™ Auto

reader 490H

clock and audible alarm as desired. The button

acts as an

Enter

 key and moves the

flashing cursor to the next changeable setting.

Setting the 3M

Attest

 Auto

reader 490H Clock

The 3M™ Attest™ Auto

reader 490H date and time are displayed in this format:

YYYY

MM

DD HH:MM, with the time displayed using a 24 hour clock. To change the

year, if necessary, push the up or down buttons to set desired year, then press

to

advance to the month, adjust as necessary and press

to advance to the day. Use the

button to move to the time and set the hours and minutes.

Setting the Audible Alarm Function

Move across the top row of text by pushing the

button. The flashing

B

 in Buzzer

indicates this is a changeable setting. It is recommended this be set at

Buzzer Enabled

to audibly alert the operator to a positive BI but it may be set to

Buzzer Disabled

 by

using either the up or down button.

Note:

 If Buzzer Disabled is selected, and the change is

accepted, the speaker icon (  ) will disappear from the LCD panel and the alarm will be

disabled for all positive 3M™ Attest™ Rapid Readout Biological Indicators, error codes,

and caution code alarm warnings.

Note:

 The 3M™ Attest™ Auto

reader 490H display backlight flashes whenever an

audible alarm sounds. Disabling the audible alarm does NOT disable the LCD display

backlight flashing.

Press

to select the desired option and advance to the Accept screen.

Accept Screen

Use the up or down buttons to toggle between Yes and No, press

 to accept changes

(Yes selected) or retain current settings (No selected) and exit the configuration mode. The

LCD panel will return to the standard display.

Summary of Contents for Attest Auto-reader 490H

Page 1: ...io 5Manual del operador 6Bedieningshandleiding 7Användarhandbok 8Brugsvejledning 9Brukerhåndbok 0Käyttöopas Manual do Utilizador Manual do Operador Εγχειρίδιο Χειριστή HInstrukcja obsługi FPříručka pro obsluhu SPriročnik za uporabo MИнструкция по эксплуатации Upute za korištenje dOperatör Kılavuzu I操作指南 jคู มือปฏิบัติงาน T작업자 매뉴얼 ...

Page 2: ......

Page 3: ...2 Product Description 2 Warranty 2 Safety and Precautions 2 Health and Safety Information 3 Explanation of Product and Packaging Label Symbols 3 Specifications 4 Instructions for Use 4 Features 7 Care and Cleaning 8 Troubleshooting Guide 8 Optional 3M Attest Auto reader 490H WebApp 9 Troubleshooting the 3M Attest Auto reader 490H WebApp 12 Repair and Replacement 12 ...

Page 4: ...fluorescence reading or a positive visual pH color change result indicate a sterilization process failure However due to the high sensitivity of the fluorescent result there is no advantage to continuing to incubate processed 3M Attest Rapid Readout Biological Indicators 1295 once the fluorescent result has been documented Other biological indicators BIs are not compatible with this device and can...

Page 5: ...H complies with the CE mark related to the Low Voltage Directive LVD 2006 95 EC as confirmed in the Declaration of Conformity The 3M Attest Auto reader 490H complies with the RoHS Directive Directive 2011 65 EU of the European Parliament and of the Council of 8 June 2011 on the restriction of the use of certain hazardous substances in electrical and electronic equipment The 3M Attest Auto reader 4...

Page 6: ...sed to sunlight or strong incandescent light Do not place instrument close to any device that emits a strong electro magnetic field Do not use on a vibrating surface Select Install Correct Power Plug The 3M Attest Auto reader 490H power supply comes with multiple blade configurations Install the blade configuration appropriate for your country into the power supply and plug into appropriate AC gro...

Page 7: ...etting Setting the 3M Attest Auto reader 490H Clock The 3M Attest Auto reader 490H date and time are displayed in this format YYYY MM DD HH MM with the time displayed using a 24 hour clock To change the year if necessary push the up or down buttons to set desired year then press to advance to the month adjust as necessary and press to advance to the day Use the button to move to the time and set t...

Page 8: ...ce the result is determined it is continuously displayed on the LCD panel until 10 seconds after the BI has been removed from the incubation well Note After the first minute of incubation if 3M Attest Rapid Readout Biological Indicator is inadvertently removed before the incubation period is complete an alarm will sound and a C2 Caution code will be displayed on the LCD panel under that well You m...

Page 9: ... positive 3M Attest Rapid Readout Biological Indicator results subsequently detected in other wells will result in the alarm sounding Removing the 3M Attest Rapid Readout Biological Indicator will reactivate the alarm for that well Pressing the Buzzer Alarm OFF button will not silence the alarm for error or caution code alarm warnings To disable the alarm see configuration section Stackable To sav...

Page 10: ...unit and plug it back in to clear error If error doesn t clear call for repair or replacement E3 Memory Error 3M Attest Auto reader 490H unable to save read data from memory Unplug the unit and plug it back in to clear error If error doesn t clear call for repair or replacement E7 Communications Error This error will occur if internal communications have failed Unplug the unit and plug it back in ...

Page 11: ...ddress URL Note To access the 3M Attest Auto reader 490H Web App within a secure network the PC or mobile device must have access to the secure network options include connecting via a network Ethernet cable signing in to the facility s wireless network or connecting via a virtual private network VPN A Serial number URL 1 Connect the supplied Ethernet cable from the 3M Attest Auto reader 490H to a...

Page 12: ...er 490H Web App Status History 3M Attest Auto reader 490H ATTEST 1295 ATTEST 1295 Auto reader 490H 1 2 3 4 5 24 6 7 8 9 10 SN 200299 2017 05 11 17 13 1 Open the Results Detail page After inserting a new 3M Attest Rapid Readout BI in the 3M Attest Auto reader 490H it is a best practice to open the 3M Attest Auto reader 490H Web App Results Detail page for that BI by clicking on the appropriate well...

Page 13: ...k on the desired radio buttons to change settings Language Selected setting indicates language in which text on 3M Attest Auto reader 490H Web App pages will appear on the PC mobile device screen Pushbutton Configuration Selecting Enable allows the user to configure the 3M Attest Auto reader 490H from the buttons on the unit itself Selecting Disable prevents the user from being able to configure t...

Page 14: ...cility virtual private network VPN on the mobile device Data entered by user on Results Detail page disappears User entered data is written to temporary web browser storage Browser may have cleared data Set web browser to accept allow or enable cookies As a best practice print or save Results Detail record as soon as BI result is final BI result displayed on 3M Attest Auto reader 490H LCD display ...

Page 15: ......

Page 16: ...t Paul MN 55144 1 800 228 3957 3M com Medical 3M Deutschland GmbH Health Care Business Carl Schurz Str 1 41453 Neuss Germany 2017 3M All rights reserved 3M and Attest are trademarks of 3M Used under license in Canada Issue Date 2017 03 34 8720 4946 4 ...

Reviews: