INSTRUCTIONS FOR USE

 READOUT PROCEDURE

Instructions for Biological Indicator (BI) Activation and Incubation

CAUTION:

To avoid the possibility of the 3M™ Attest™ Rapid Readout

Biological Indicator 1295 vial absorbing fluorescent residue from a chemical indicator

or tape, place the 3M™ Attest™ Rapid Readout Biological Indicator vial so it does not

come in direct contact with chemical indicators or tape.

WARNING:

To reduce the risk associated with incorrect results

Do not incubate a 3M™ Attest™ Rapid Readout Biological Indicator 1295 if, after

processing and before BI activation, it is observed to have a broken media ampoule.

Retest the sterilizer with a new biological indicator.

Activate and incubate 3M™ Attest™ Rapid Readout Biological Indicators 1295 within

1 hour of the completion of the sterilization cycle.

Incubate the 3M™ Attest™ Rapid Readout Biological Indicator only in the

3M™ Attest™ Auto

reader 490H i.e., match the pink cap color of the 1295 BI with

the pink color

coded configuration sticker surrounding the 3M™ Attest™ Auto

reader

490H incubation wells.

CAUTION:

Wear safety glasses and gloves when activating the 3M™ Attest™

Rapid Readout Biological Indicator 1295.

Follow this procedure for every 3M™ Attest™ Rapid Readout Biological Indicator (BI) to

be read.

or

While wearing safety glasses and gloves,

PLACE

BI in color

coded 3M™ Attest™

Auto

reader 490H well. Note that BI

vial is keyed for specific orientation

in well (flat back of BI faces back of

3M™ Attest™ Auto

reader 490H).

While wearing safety glasses and gloves,

PLACE

 BI in Attest™ Biological Indicator

Activator.

or

CLOSE

 BI cap and

CRUSH

 media

ampoule by firmly pushing vial straight

down in incubation well.

CLOSE

 and

SQUEEZE

 Attest™ Biological

Indicator Activator to close BI cap and crush

media ampoule.

Immediately remove BI, visually

VERIFY

 the glass ampoule

(containing growth media) has been crushed, and

FLICK

 BI to

ensure purple growth media flows to the growth chamber at the

bottom of the vial.

PLACE

 the activated BI into a 3M™ Attest™ Auto-reader 490H

well and observe 3M™ Attest™ Auto-reader 490H LCD display

to confirm correct placement of the BI. If the BI is correctly

seated, the 3M™ Attest™ Auto-reader 490H LCD display will

show ***, **, * and then remaining minutes of incubation directly

below the incubating BI.

RECORD

 BI result.

WARNING:

 Do not remove the 3M™ Attest™ Rapid Readout Biological

Indicator 1295 from the incubation well until the (+) or (

) symbol on the LCD display

indicates the test is complete.

INSTRUCTIONS FOR USE

 ADDITIONAL INFORMATION

BI Placement

Remaining incubation time, in minutes, will be displayed on the LCD display to indicate

the 3M™ Attest™ Rapid Readout Biological Indicator is incubating in the reader well and

should not be touched. The time remaining display indicates the result is not available and

incubation is in progress.

Do not remove 3M™ Attest™ Rapid Readout Biological Indicator until the (+) or (-)

symbol on the LCD panel indicates the test is complete.

 Once the result is determined,

it is continuously displayed on the LCD panel until 10 seconds after the BI has been

removed from the incubation well.

Note:

After the first minute of incubation, if 3M™ Attest™ Rapid Readout Biological

Indicator is inadvertently removed before the incubation period is complete, an alarm will

sound and a C2 Caution code will be displayed on the LCD panel under that well.

You

must return the 3M™ Attest™ Rapid Readout Biological Indicator to the reader well

within 10 seconds or the results will be lost.

WARNING:

 Do not remove or change placement of 3M™ Attest™ Rapid

Readout Biological Indicator once it is placed into a well. Doing so may result in loss of

results and the 3M™ Attest™ Rapid Readout Biological Indicator test may be invalid.

INSTRUCTIONS FOR USE

 POSITIVE CONTROL

Positive Control

Activate and incubate at least one non

processed 3M™ Attest™ Rapid Readout Biological

Indicator 1295 to use as the positive control each day that a processed 3M™ Attest™

Rapid Readout Biological Indicator 1295 is incubated and read by the unit. Write a

C

 (for

control

) and the date on the label. The positive control can be placed in any incubation

well. The positive control should be the same part number and have the same lot number

as the processed 3M™ Attest™ Rapid Readout Biological Indicator in the unit.
The positive control biological indicator helps confirm:
• correct incubation temperatures are met (heater block reading of 60°C +/

2°C);

• viability of spores has not been altered due to improper storage temperature, humidity,

or proximity to chemicals;

• capability of media to promote rapid growth; and

• proper functioning of 3M™ Attest™ Auto

reader 490H components.

INSTRUCTIONS FOR USE

 INTERPRETATION OF RESULTS

Fluorescent Results

After placing a 3M™ Attest™ Rapid Readout Biological Indicator into an incubation

well, the 3M™ Attest™ Auto

reader 490H LCD display will show ***, **, * and then

the remaining minutes of incubation directly below the incubating BI to indicate that

incubation/reading is in progress. The 3M™ Attest™ Rapid Readout Biological Indicator

should not be removed or touched when the display indicates remaining incubation time.
A (+) will display on the LCD panel directly under the BI and an alarm will sound (if the

sound feature is enabled) as soon as a positive 3M™ Attest™ Rapid Readout Biological

Indicator result is detected.
At the end of the specified incubation time, if a negative 3M™ Attest™ Rapid Readout

Biological Indicator result is detected, a (

) will display on the LCD panel directly under

the incubating BI. The final negative fluorescent result (

) for a processed 3M™ Attest™

Rapid Readout Biological Indicator 1295 indicates an acceptable sterilization process.
After either the (+) or (

) symbol is displayed, and the result is recorded, the

3M™ Attest™ Rapid Readout Biological Indicator can be removed from the well. Do not

re-incubate BIs as the result will not be valid.
The positive (unprocessed) control 3M™ Attest™ Rapid Readout Biological Indicator

1295 must provide a positive fluorescent result (+). Processed 3M™ Attest™ Rapid

Readout Biological Indicator results are not valid until the positive control reads fluorescent

positive (+).

Summary of Contents for Attest Auto-reader 490H

Page 1: ...io 5Manual del operador 6Bedieningshandleiding 7Användarhandbok 8Brugsvejledning 9Brukerhåndbok 0Käyttöopas Manual do Utilizador Manual do Operador Εγχειρίδιο Χειριστή HInstrukcja obsługi FPříručka pro obsluhu SPriročnik za uporabo MИнструкция по эксплуатации Upute za korištenje dOperatör Kılavuzu I操作指南 jคู มือปฏิบัติงาน T작업자 매뉴얼 ...

Page 2: ......

Page 3: ...2 Product Description 2 Warranty 2 Safety and Precautions 2 Health and Safety Information 3 Explanation of Product and Packaging Label Symbols 3 Specifications 4 Instructions for Use 4 Features 7 Care and Cleaning 8 Troubleshooting Guide 8 Optional 3M Attest Auto reader 490H WebApp 9 Troubleshooting the 3M Attest Auto reader 490H WebApp 12 Repair and Replacement 12 ...

Page 4: ...fluorescence reading or a positive visual pH color change result indicate a sterilization process failure However due to the high sensitivity of the fluorescent result there is no advantage to continuing to incubate processed 3M Attest Rapid Readout Biological Indicators 1295 once the fluorescent result has been documented Other biological indicators BIs are not compatible with this device and can...

Page 5: ...H complies with the CE mark related to the Low Voltage Directive LVD 2006 95 EC as confirmed in the Declaration of Conformity The 3M Attest Auto reader 490H complies with the RoHS Directive Directive 2011 65 EU of the European Parliament and of the Council of 8 June 2011 on the restriction of the use of certain hazardous substances in electrical and electronic equipment The 3M Attest Auto reader 4...

Page 6: ...sed to sunlight or strong incandescent light Do not place instrument close to any device that emits a strong electro magnetic field Do not use on a vibrating surface Select Install Correct Power Plug The 3M Attest Auto reader 490H power supply comes with multiple blade configurations Install the blade configuration appropriate for your country into the power supply and plug into appropriate AC gro...

Page 7: ...etting Setting the 3M Attest Auto reader 490H Clock The 3M Attest Auto reader 490H date and time are displayed in this format YYYY MM DD HH MM with the time displayed using a 24 hour clock To change the year if necessary push the up or down buttons to set desired year then press to advance to the month adjust as necessary and press to advance to the day Use the button to move to the time and set t...

Page 8: ...ce the result is determined it is continuously displayed on the LCD panel until 10 seconds after the BI has been removed from the incubation well Note After the first minute of incubation if 3M Attest Rapid Readout Biological Indicator is inadvertently removed before the incubation period is complete an alarm will sound and a C2 Caution code will be displayed on the LCD panel under that well You m...

Page 9: ... positive 3M Attest Rapid Readout Biological Indicator results subsequently detected in other wells will result in the alarm sounding Removing the 3M Attest Rapid Readout Biological Indicator will reactivate the alarm for that well Pressing the Buzzer Alarm OFF button will not silence the alarm for error or caution code alarm warnings To disable the alarm see configuration section Stackable To sav...

Page 10: ...unit and plug it back in to clear error If error doesn t clear call for repair or replacement E3 Memory Error 3M Attest Auto reader 490H unable to save read data from memory Unplug the unit and plug it back in to clear error If error doesn t clear call for repair or replacement E7 Communications Error This error will occur if internal communications have failed Unplug the unit and plug it back in ...

Page 11: ...ddress URL Note To access the 3M Attest Auto reader 490H Web App within a secure network the PC or mobile device must have access to the secure network options include connecting via a network Ethernet cable signing in to the facility s wireless network or connecting via a virtual private network VPN A Serial number URL 1 Connect the supplied Ethernet cable from the 3M Attest Auto reader 490H to a...

Page 12: ...er 490H Web App Status History 3M Attest Auto reader 490H ATTEST 1295 ATTEST 1295 Auto reader 490H 1 2 3 4 5 24 6 7 8 9 10 SN 200299 2017 05 11 17 13 1 Open the Results Detail page After inserting a new 3M Attest Rapid Readout BI in the 3M Attest Auto reader 490H it is a best practice to open the 3M Attest Auto reader 490H Web App Results Detail page for that BI by clicking on the appropriate well...

Page 13: ...k on the desired radio buttons to change settings Language Selected setting indicates language in which text on 3M Attest Auto reader 490H Web App pages will appear on the PC mobile device screen Pushbutton Configuration Selecting Enable allows the user to configure the 3M Attest Auto reader 490H from the buttons on the unit itself Selecting Disable prevents the user from being able to configure t...

Page 14: ...cility virtual private network VPN on the mobile device Data entered by user on Results Detail page disappears User entered data is written to temporary web browser storage Browser may have cleared data Set web browser to accept allow or enable cookies As a best practice print or save Results Detail record as soon as BI result is final BI result displayed on 3M Attest Auto reader 490H LCD display ...

Page 15: ......

Page 16: ...t Paul MN 55144 1 800 228 3957 3M com Medical 3M Deutschland GmbH Health Care Business Carl Schurz Str 1 41453 Neuss Germany 2017 3M All rights reserved 3M and Attest are trademarks of 3M Used under license in Canada Issue Date 2017 03 34 8720 4946 4 ...

Reviews: