•  For Extended Mode, an acoustical signal is emitted 

every 3 seconds for a total of 15 seconds.

•  If desired, press the START button to stop the light 

emission before the set time is over.

Tack Cure Function

Hold down the START button to activate the Tack Cure 

function regardless of the set Exposure Time. The unit 

emits a single one-second light pulse which allows for 

a short exposure to tack cure brackets. The light will 

return to standard operation when the START button is 

no longer depressed.

Recommended Cure Techniques with  

3M Unitek Adhesives

•    Rotate the light guide into the desired position  

for polymerization.

•     For most 3M Unitek adhesives, reference the table 

below for recommended cure technique. 

•  For other 3M Unitek adhesives not specifically 

mentioned, please refer to the adhesive Instructions 

for Use.

•   If using other light cure adhesive systems, consult 

the adhesive manufacturer’s instructions.

•   The light guide tip should be placed approximately  

1-2 mm above the appliance and pointed 

perpendicular towards the tooth surface.

Operation

Activating the Handpiece

If the Power Level LED is off, briefly press the START 

button to activate the hand piece. The Power Level LED 

will come on and a short audible signal is emitted. The 

hand piece is now ready for operation; the handpiece 

displays the latest set exposure time setting.

Selection of Exposure Time

Press the TIME button to cycle through the following 

settings from left to right. Press the TIME button again 

to go back to 3s settings from the Extended Mode.

Caution!

 High intensity light always involves the 

production of heat. When utilizing the light in  

Extended Mode (15s), do not keep the light in  

one position. Longer exposure in the region of the pulp 

can lead to injury or irritation of the pulp, which is why 

the specified exposure times must not be exceeded.

Activating and Deactivating the  

Light Emission

Briefly press the START button to activate the light 

emission for the Exposure Time set as described above. 
•  Once the emission has begun, the active Exposure 

Time LEDs will count down every 3s with an acoustical 

signal indicating the elapsed time, except for the 

Extended Mode. When the emission is completed, the 

Exposure Time LEDs will return as set before. 

•  Example with 12s Exposure Time setting:

START button

Power Level LED

TIME button

LEDs showing 

exposure times

3s

6s

9s

12s

Extended  

Mode (15s)

Beginning 

of Emission

3 sec 

Elapsed

6 sec 

Elapsed

9 sec 

Elapsed

End of 

Emission

ON Signal

1 beep

2 beeps

3 beeps OFF Signal

Appliance

Adhesive

Ortholux

 

Luminous 

Curing Light 

Metal  

Brackets

Transbond

 XT Adhesive,  

Transbond

 PLUS Color 

Change Adhesive,  

APC

 II Adhesive  

Coated Brackets,  

APC

 PLUS Adhesive 

Coated Brackets

3 seconds mesial x 

3 seconds distal

Ceramic  

Brackets

3 seconds through  

the bracket

Buccal Tubes  

(direct bond)

6

 

seconds mesial x 

6 seconds occlusal

Lingual  

Retainers

Transbond

 LR Adhesive

3 seconds mesial x 

3 seconds distal

Molar  

Bands

Transbond

 Plus  

Band Adhesive

12 seconds  

(3 seconds  

per cusp)

Unitek

 Multi-Cure Glass 

Ionomer Band Cement

Indirect Trays  

(Metal Brackets)

Transbond

 Supreme  

LV Adhesive

6 seconds mesial 

x 6 seconds distal 

through the tray

Indirect Trays  

(Ceramic 

Brackets)

6 seconds through  

the bracket  

and tray

Summary of Contents for Ortholux Luminous Curing Ligh

Page 1: ...ENGLISH Fast Curing Cordless LED Light Instructions For Use Ortholux Luminous Curing Light...

Page 2: ...age stated on the rating plate is compatible with the available main voltage Operation of the unit at a different voltage may damage the unit 3 Position the unit so that the power plug is accessible a...

Page 3: ...devices 18 Do not use Ortholux Luminous Curing Light for patients with a history of photobiological reactions including individuals with urticaria solaris or erythropoetic protoporphyria or who are c...

Page 4: ...The handpiece switches to sleep mode once it is placed in the charger or if left unused for approximately 5 minutes outside the charger NOTE Do not discard these Instructions for the duration of produ...

Page 5: ...o technical modification without prior notice Installation of the Unit Factory Settings The factory settings of the unit are as follows When first turned on the unit is in Extended Mode i e only the P...

Page 6: ...ge the new battery completely for the first time The status LED showing the operating status of the charger flashes green during the charging operation see below Operating Status Display of the Charge...

Page 7: ...om the Extended Mode Caution High intensity light always involves the production of heat When utilizing the light in Extended Mode 15s do not keep the light in one position Longer exposure in the regi...

Page 8: ...varying arrangement of the components of the units NOTE The light intensity is independent of the battery charge level Caution Clean the testing area with a damp cloth Without applying pressure place...

Page 9: ...y after the acoustical signal has gone off 2 minute cooling off time in the air typically corresponds to 1 minute operating time The light emission does not start when the START button is pressed an e...

Page 10: ...defective batteries and units in accordance with local legal regulations Maintenance and Care The Ortholux Luminous Curing Light is maintenance free with the exception of cleaning the light guide and...

Page 11: ...A PARTICULAR PURPOSE User is responsible for determining the suitability of the product for user s application If this product is defective within the warranty period your exclusive remedy and 3M Unit...

Page 12: ...chien Tel 00420 47 5620808 Fax 00420 47 5620299 DSD Dental Service Dienst Jaap Wim Hr John Wolters de Jong Bosma Ketting Markerkant 13 03 i NL 1314 AL Amere Netherlands Tel 0031 36 53 4 32 12 Fax 0031...

Page 13: ...oup 1 The Ortholux Luminous Curing Light unit uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic eq...

Page 14: ...e 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on...

Page 15: ...determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol N...

Page 16: ...eparation distance according to frequency of transmitter m Rated maximum output of transmitter W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 04 0 12 0 40 0 1 0 11 0 38 1 26 1 0 35 1...

Reviews: