• For Extended Mode, an acoustical signal is emitted
every 3 seconds for a total of 15 seconds.
• If desired, press the START button to stop the light
emission before the set time is over.
Tack Cure Function
Hold down the START button to activate the Tack Cure
function regardless of the set Exposure Time. The unit
emits a single one-second light pulse which allows for
a short exposure to tack cure brackets. The light will
return to standard operation when the START button is
no longer depressed.
Recommended Cure Techniques with
3M Unitek Adhesives
• Rotate the light guide into the desired position
for polymerization.
• For most 3M Unitek adhesives, reference the table
below for recommended cure technique.
• For other 3M Unitek adhesives not specifically
mentioned, please refer to the adhesive Instructions
for Use.
• If using other light cure adhesive systems, consult
the adhesive manufacturer’s instructions.
• The light guide tip should be placed approximately
1-2 mm above the appliance and pointed
perpendicular towards the tooth surface.
Operation
Activating the Handpiece
If the Power Level LED is off, briefly press the START
button to activate the hand piece. The Power Level LED
will come on and a short audible signal is emitted. The
hand piece is now ready for operation; the handpiece
displays the latest set exposure time setting.
Selection of Exposure Time
Press the TIME button to cycle through the following
settings from left to right. Press the TIME button again
to go back to 3s settings from the Extended Mode.
Caution!
High intensity light always involves the
production of heat. When utilizing the light in
Extended Mode (15s), do not keep the light in
one position. Longer exposure in the region of the pulp
can lead to injury or irritation of the pulp, which is why
the specified exposure times must not be exceeded.
Activating and Deactivating the
Light Emission
Briefly press the START button to activate the light
emission for the Exposure Time set as described above.
• Once the emission has begun, the active Exposure
Time LEDs will count down every 3s with an acoustical
signal indicating the elapsed time, except for the
Extended Mode. When the emission is completed, the
Exposure Time LEDs will return as set before.
• Example with 12s Exposure Time setting:
START button
Power Level LED
TIME button
LEDs showing
exposure times
3s
6s
9s
12s
Extended
Mode (15s)
Beginning
of Emission
3 sec
Elapsed
6 sec
Elapsed
9 sec
Elapsed
End of
Emission
ON Signal
1 beep
2 beeps
3 beeps OFF Signal
Appliance
Adhesive
Ortholux
™
Luminous
Curing Light
Metal
Brackets
Transbond
™
XT Adhesive,
Transbond
™
PLUS Color
Change Adhesive,
APC
™
II Adhesive
Coated Brackets,
APC
™
PLUS Adhesive
Coated Brackets
3 seconds mesial x
3 seconds distal
Ceramic
Brackets
3 seconds through
the bracket
Buccal Tubes
(direct bond)
6
seconds mesial x
6 seconds occlusal
Lingual
Retainers
Transbond
™
LR Adhesive
3 seconds mesial x
3 seconds distal
Molar
Bands
Transbond
™
Plus
Band Adhesive
12 seconds
(3 seconds
per cusp)
Unitek
™
Multi-Cure Glass
Ionomer Band Cement
Indirect Trays
(Metal Brackets)
Transbond
™
Supreme
LV Adhesive
6 seconds mesial
x 6 seconds distal
through the tray
Indirect Trays
(Ceramic
Brackets)
6 seconds through
the bracket
and tray