Insert the battery:
•
Remove the protective cap from the battery
and keep with the unit packaging.
• Slowly insert the battery with the thread side
towards the metal housing into the handpiece
until it stops. Screw the battery clockwise by hand
until the seal is pressed firmly against the metal
housing. The unit will not function properly if the
battery is not screwed in completely!
• If the unit does not function properly, remove
the battery and reinsert it as described above.
Power Level Display of the Handpiece
Battery Charging
• The unit contains a powerful lithium-ion
rechargeable battery. The battery does not have any
memory effect and can therefore be recharged at
any time by placing it in the charger (see “Power
Level Display of the Handpiece” below).
• Alternatively, the battery in the unit can be replaced
by a spare battery (available separately) which has
been charged.
• Place the handpiece on the charger for approximately
1.5 hours before the first use, in order to charge the
new battery completely for the first time.
- The status LED showing the operating status of
the charger flashes green during the charging
operation; see below “Operating Status Display
of the Charger.”
Operating Status Display of the Charger
Status LED
Steady green light
Flashes green
Steady red light
Flashes alternately red and green
Handpiece/battery outside the charger
Charger ready for operation
n/a
Charge contact pins are wet
Malfunction in the charger
Handpiece/battery in the charger
Charging has been completed
Battery is being charged
Charge contact pins are wet
Problem during charging
Status LED showing
operating status
Power Level LED
Operating status
Power Level LED
Steady green light
Steady red light
Flashes red
Handpiece outside the charger
Battery is charged, handpiece ready for operation
Warning for low battery capacity, remaining capacity
adequate typically for 20 3-second cycles
Battery fully discharged, exposure cycle will be
completed or, if in extended mode, stopped
Handpiece in the charger
Not possible, handpiece is in sleep mode
Not possible, handpiece is in sleep mode
Problem in charging, battery is defective
or cannot be charged
Operating status
Seal