•   The recommendations are applicable with 100% 

light output intensity, measured at the intensity test 

area on the charging base.

•  A technique tip for faster light curing of metal 

brackets is to position the light guide interproximal 

to two brackets. In order for the bracket to fully 

cure, both sides of the bracket must be exposed to 

the light.

Keep the light guide clean at all times to 

obtain full light intensity.

 Cured primer/

adhesive on the light guide will reduce the 

intensity resulting in incomplete cure.

Damaged light guides substantially reduce 

the light intensity and must be replaced 

immediately, sharp edges may cause 

serious injury!

NOTE: Holding the light away from the tooth, moving 

the light guide and turning the light guide tip at an 

angle may cause incomplete cure of the material.

Sleep Mode

Once the handpiece is placed in the charger, all 

internal functions and LEDs are automatically turned 

off as the handpiece switches to sleep mode. This 

reduces the power consumption of the battery to a 

minimal level. Outside the charger, the handpiece 

is also switched to sleep mode if it is not used 

for approximately 5 minutes. See “Activating the 

Handpiece” to wake up the unit.

Measurement of Light Intensity

The light intensity of the Ortholux Luminous Curing 

Light handpiece can only be reliably determined 

using the light intensity testing area on the Ortholux 

Luminous Curing Light charger! The circular testing 

area is located below the Status LED on the charger. 

Measurements carried out using other units will 

produce incorrect results due to the different 

light sources and the varying arrangement of the 

components of the units.

NOTE: The light intensity is independent of the 

battery charge level.

Caution!
•  Clean the testing area with a damp cloth.
•  Without applying pressure, place the light guide 

on the testing area such that the light-emitting 

opening of the light guide is flush with the 

testing area.

•  Activate the lamp by pressing the START button.

-  The number of illuminated LEDs on the 

charger is indicative of the measured light 

intensity: 5 LEDs = 100%, 4 LEDs = 90%, 3 

LEDs = 80%, 2 LEDs = 70%, 1 LED = 60%.

• 

If the light intensity is below 80% (fewer 

than 3 LEDs are ON), check the light guide 

and handpiece for contamination or defects.

•  Remove and clean the light guide of any 

contamination. Replace a defective light guide 

with a new light guide. If the measures described 

above do not result in any improvement, call  

3M Unitek Customer Service.

Acoustical Signals - Handpiece

An acoustical signal is emitted
•  every time a button is pressed
•  every time the light is turned ON or OFF
•  1 time after 3 seconds exposure time, 2 times 

after 6 seconds, 3 times after 9 seconds 

Exception: in Extended Mode; an acoustical 

signal is emitted every 3 seconds

•  every time the sleep mode is terminated by 

pressing the START button.

A 2 second-error signal is emitted, if
•  the handpiece overheats
•  the battery lacks sufficient charge

The acoustical signals from the handpiece can  

be turned off (except for the 2-second error signal). 

Put the handpiece in sleep, e.g., by setting it in the 

charger. Take the device from the charger; press 

the TIME button, then press the START button. This 

changes the sleep mode of the handpiece back to 

active and the status “Acoustical signals activated” 

to “Acoustical signals deactivated”. The acoustical 

signals can be reactivated by following the  

same procedure. 

Testing area

LEDs showing 

light intensity

Summary of Contents for Ortholux Luminous Curing Ligh

Page 1: ...ENGLISH Fast Curing Cordless LED Light Instructions For Use Ortholux Luminous Curing Light...

Page 2: ...age stated on the rating plate is compatible with the available main voltage Operation of the unit at a different voltage may damage the unit 3 Position the unit so that the power plug is accessible a...

Page 3: ...devices 18 Do not use Ortholux Luminous Curing Light for patients with a history of photobiological reactions including individuals with urticaria solaris or erythropoetic protoporphyria or who are c...

Page 4: ...The handpiece switches to sleep mode once it is placed in the charger or if left unused for approximately 5 minutes outside the charger NOTE Do not discard these Instructions for the duration of produ...

Page 5: ...o technical modification without prior notice Installation of the Unit Factory Settings The factory settings of the unit are as follows When first turned on the unit is in Extended Mode i e only the P...

Page 6: ...ge the new battery completely for the first time The status LED showing the operating status of the charger flashes green during the charging operation see below Operating Status Display of the Charge...

Page 7: ...om the Extended Mode Caution High intensity light always involves the production of heat When utilizing the light in Extended Mode 15s do not keep the light in one position Longer exposure in the regi...

Page 8: ...varying arrangement of the components of the units NOTE The light intensity is independent of the battery charge level Caution Clean the testing area with a damp cloth Without applying pressure place...

Page 9: ...y after the acoustical signal has gone off 2 minute cooling off time in the air typically corresponds to 1 minute operating time The light emission does not start when the START button is pressed an e...

Page 10: ...defective batteries and units in accordance with local legal regulations Maintenance and Care The Ortholux Luminous Curing Light is maintenance free with the exception of cleaning the light guide and...

Page 11: ...A PARTICULAR PURPOSE User is responsible for determining the suitability of the product for user s application If this product is defective within the warranty period your exclusive remedy and 3M Unit...

Page 12: ...chien Tel 00420 47 5620808 Fax 00420 47 5620299 DSD Dental Service Dienst Jaap Wim Hr John Wolters de Jong Bosma Ketting Markerkant 13 03 i NL 1314 AL Amere Netherlands Tel 0031 36 53 4 32 12 Fax 0031...

Page 13: ...oup 1 The Ortholux Luminous Curing Light unit uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic eq...

Page 14: ...e 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on...

Page 15: ...determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol N...

Page 16: ...eparation distance according to frequency of transmitter m Rated maximum output of transmitter W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 04 0 12 0 40 0 1 0 11 0 38 1 26 1 0 35 1...

Reviews: