8.  Ortholux Luminous Curing Light may be operated 

only with the supplied light guide or 3M Unitek 

replacement light guide. The use of other light 

guides may result in a reduction or increase in the 

light intensity. The product’s warranty does not 

cover any damage resulting from the use of third-

party light guides.

9.   Condensation resulting from the unit being 

transferred from a cold to a warm environment may 

be a potential risk. Hence, the unit should be turned 

on only after it is completely equilibrated to ambient 

temperature.

10. In order to avoid electric shock do not introduce 

any objects into the unit with the exception of 

replacement parts handled in accordance with the 

Instructions for Use.

11. Use only genuine 3M Unitek parts when replacing 

defective components as directed in these 

Instructions for Use. The product’s warranty does 

not cover any damage resulting from the use of 

third-party replacement parts. 

12. Should you have any reason to suspect the safety 

of the unit to be compromised, the unit must be 

taken out of operation and labeled accordingly 

to prevent third parties from inadvertently 

using a possibly defective unit. Safety may be 

compromised, e.g., if the unit malfunctions or is 

noticeably damaged.

13. Keep solvents, flammable liquids, and sources 

of intense heat away from the unit as they may 

damage the plastic housing of the charger, the 

seals, or the cover on the operating buttons.

14. Do not operate the unit in the proximity of 

flammable mixtures.

15. Do not allow cleaning agents to enter the unit as 

this may short-circuit the unit or cause potentially 

dangerous malfunction.

16. Only service centers authorized by 3M Unitek may 

open the unit housing and repair the device.

17. Ortholux Luminous Curing Light must not be used 

for patients, or by users, with heart pacemaker 

implants who have been advised to be cautious with 

regard to their exposure to small electrical devices.

18. Do not use Ortholux Luminous Curing Light 

for patients with a history of photobiological 

reactions (including individuals with urticaria 

solaris or erythropoetic protoporphyria) or who are 

currently on photosensitizing medication (including 

8-methoxypsoralen or dimethylchlorotetracycline).

19. Individuals with a history of cataract surgery may 

be particularly sensitive to the exposure to light 

and should be discouraged from exposure to the 

Ortholux Luminous Curing Light unless adequate 

safety measures, such as the use of protective 

goggles to filter blue light, are undertaken.

20. Individuals with a history of retinal disease should 

seek advice from their ophthalmologist prior 

to operating the unit. In operating the Ortholux 

Luminous Curing Light, this group of individuals 

must take extreme care and comply with any and 

all safety precautions (including the use of suitable 

light-filtering safety goggles).

21. This unit has been developed and tested in 

accordance with the relevant EMC regulations 

and standards. It is in conformity with legal 

requirements. Since various factors such as power 

supply, wiring, and the ambient conditions at the 

place of operation can affect the EMC properties 

of the unit, the possibility that, under unfavorable 

conditions, there will be EMC disruptions cannot be 

completely excluded. If you should notice problems 

in the operation of this or other units, move the unit 

to a different location.

22. Prior to each use of the unit ensure that the emitted 

light intensity is sufficient to safely guarantee 

polymerization. Use the light testing area integrated 

into the charger for this purpose.

Glossary of Symbols

Attention, Consult Accompanying 

Documents

Type B Equipment -  

Protection against electric shock

Class II Equipment - double insulated

Establishes that the product conforms to all 

applicable legal requirements

NRTL approval applies only to  

USA version PN 704-460

Battery power level

Icon to identify electric and electronic 

devices. The unit must be collected and 

disposed of separately.

Temperature limitation

Humidity limitation

Atmospheric pressure limitation

Summary of Contents for Ortholux Luminous Curing Ligh

Page 1: ...ENGLISH Fast Curing Cordless LED Light Instructions For Use Ortholux Luminous Curing Light...

Page 2: ...age stated on the rating plate is compatible with the available main voltage Operation of the unit at a different voltage may damage the unit 3 Position the unit so that the power plug is accessible a...

Page 3: ...devices 18 Do not use Ortholux Luminous Curing Light for patients with a history of photobiological reactions including individuals with urticaria solaris or erythropoetic protoporphyria or who are c...

Page 4: ...The handpiece switches to sleep mode once it is placed in the charger or if left unused for approximately 5 minutes outside the charger NOTE Do not discard these Instructions for the duration of produ...

Page 5: ...o technical modification without prior notice Installation of the Unit Factory Settings The factory settings of the unit are as follows When first turned on the unit is in Extended Mode i e only the P...

Page 6: ...ge the new battery completely for the first time The status LED showing the operating status of the charger flashes green during the charging operation see below Operating Status Display of the Charge...

Page 7: ...om the Extended Mode Caution High intensity light always involves the production of heat When utilizing the light in Extended Mode 15s do not keep the light in one position Longer exposure in the regi...

Page 8: ...varying arrangement of the components of the units NOTE The light intensity is independent of the battery charge level Caution Clean the testing area with a damp cloth Without applying pressure place...

Page 9: ...y after the acoustical signal has gone off 2 minute cooling off time in the air typically corresponds to 1 minute operating time The light emission does not start when the START button is pressed an e...

Page 10: ...defective batteries and units in accordance with local legal regulations Maintenance and Care The Ortholux Luminous Curing Light is maintenance free with the exception of cleaning the light guide and...

Page 11: ...A PARTICULAR PURPOSE User is responsible for determining the suitability of the product for user s application If this product is defective within the warranty period your exclusive remedy and 3M Unit...

Page 12: ...chien Tel 00420 47 5620808 Fax 00420 47 5620299 DSD Dental Service Dienst Jaap Wim Hr John Wolters de Jong Bosma Ketting Markerkant 13 03 i NL 1314 AL Amere Netherlands Tel 0031 36 53 4 32 12 Fax 0031...

Page 13: ...oup 1 The Ortholux Luminous Curing Light unit uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic eq...

Page 14: ...e 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on...

Page 15: ...determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol N...

Page 16: ...eparation distance according to frequency of transmitter m Rated maximum output of transmitter W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 04 0 12 0 40 0 1 0 11 0 38 1 26 1 0 35 1...

Reviews: