![background image](http://html1.mh-extra.com/html/3dhistech/de2bghm00010100001/de2bghm00010100001_user-manual_442155094.webp)
PANNORAMIC® DESK DW II 3.0.4
SETTINGS AND SERVICE
MARCH 30, 2021
93
▪
Off
: a value of 1.0 is used.
o
Current threshold
–
Set the desired threshold level in the box.
▪
Use selection box
–
If selected, only the area inside the box is scanned. It
is useful if you only want to scan the area that is equivalent with the size
of the coverslip, so external specks will not be scanned.
o
Stitching overlap
–
Set the amount how much the FOVs should extend over.
o
Measure camera rotation
–
Camera rotation angle is measured.
o
Lower image position
–
Moves the upper image sideways in distinctive steps set
in Step size value box.
o
Reset all camera rotation
–
Click this button to set camera rotation values to
default.
Save settings by clicking
.
•
–
Image compensation
You can use a saved compensation image whenever compensation-related issues occur
(deriving from the slide, illumination, staining or embedding medium irregularities).
Figure 79 Image compensation window
o
BF cameras
field displays the active camera. Currently, there is only one active
camera for Brightfield mode.
o
Objectives
field displays the objective used for scanning.
There might be specks or minor etches on the surface of the compensation slide that
require modifying the threshold settings. Therefore, the scanner will not use those areas
for the creation of the compensation images. To modify the threshold settings, set the
desired threshold value in the box.