![background image](http://html1.mh-extra.com/html/3dhistech/de2bghm00010100001/de2bghm00010100001_user-manual_442155054.webp)
PANNORAMIC® DESK DW II 3.0.4
SCANNING MODES AND SETTINGS
MARCH 30, 2021
53
Figure 36 – Focus mode
Sample detection
This panel offers options for threshold setting and selection of the area to be scanned.
Figure 37 – Sample detection window
•
Source
o
Brightfield preview
–
a normal type of illumination in the microscope for scanning
the slide. This option is recommended when scanning in Brightfield mode.
•
Thresholding
The manually specified value is used for thresholding during scanning. Preview image of
the loaded slide is displayed in the Preview window, and when modifying any of the
parameter values, it will have effect on the preview image. If the threshold value
presented on the Microscope Controller preview window is modified, this value is
automatically applied by the profile in the Profile Settings panel.