PANNORAMIC® DESK DW II 3.0.4
SCANNING MODES AND SETTINGS
MARCH 30, 2021
61
Caution!
It requires lots of time (since it cannot find focus neither on the edges, nor on specks) to
digitize coverslip edges. Select
Remove coverslip
function to eliminate those areas.
Warning!
Do
NOT
use
Remove coverslip
function on TMA slides! The layout of TMA slides may mislead
the detecting algorithm.
Notice:
If a hematology slide is detected including blood smear, the
Remove Coverslip
function
cannot be activated and focusing on the coverslip area is not available. If you want to use
the
Remove Coverslip
function, first switch to
Custom Scanmap
.
Fill holes inside scanmap
–
If selected, the holes inside the specimen will also be
scanned. Often these holes contain small sample objects that otherwise would not be
scanned.
Remove specks
–
If selected, the scanner does not digitize objects smaller than the
predetermined size. This helps filtering specks and other uninteresting objects out.
Set the maximum volume of specks manually in the setting window
. The maximum
value that can be set is 1.