Appendix A
37
/
44
closer
than
30
cm
(12
inches)
to
any
part
of
theNAVI-30/NAVI-60 vein illuminator, including cables
specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
Guidance and manufacturer’s
declaration
–
electromagnetic emissions
The
NAVI-30/NAVI-60 vein illuminator
is intended for use in the
electromagnetic environment specified below.
The customer or the user of the NAVI-30/NAVI-60 vein illuminator
should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment
–
guidance
RF emission
CISPR 11
Group 1
The NAVI-30/NAVI-60 vein
illuminator
uses RF energy only for
its internal function. Therefore, its
RF emissions are very low and are
not likely to cause any interference
in nearby electronic equipment.
Radio-frequency
emission
CISPR 11
Class A
The NAVI-30/NAVI-60 vein
illuminator
is suitable for use in all
establishments other than domestic
and those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
Harmonic
emission
IEC61000-3-2
Class A
Voltage
fluctuation and
flashing
IEC 61000-3-3
Complies