Common Faults
25
/
44
Solution
: Connect the
device
to the power adapter for charging
the battery (the battery indicator is extinguished when the
battery is fully charged), or directly use the external power
supply to power the device. If the device still cannot be powered
on, it may be abnormal. In this case, contact the after-sales
service department of MEDCAPTAIN.
The continuous operation duration of the battery is
insufficient.
Possible cause
: The battery is not fully charged, or the
continuous operation duration of the battery decreases due to
natural deterioration.
Check method
: Connect the device to the power adapter. If the
battery indicator is blinking, the battery is not fully charged. If
the battery indicator is not illuminated, the battery is already
fully charged. The decrease in the continuous operation duration
of the battery is caused by increase in number of
charge/discharge times, which is natural and inevitable.
Solution
: If the continuous operation duration of the battery is
too short, the battery needs to be replaced. Please consult the
after-sales service department of MEDCAPTAIN about battery
replacement.
The device does not respond or breaks down occasionally.
Possible
cause: The device has been used for a long time and the
internal temperature is too high.
Solution
: Power off the device for a moment to cool down the
device.