Advanced programming
ZOLL Lifecor internal use only
Page B-11
Treatment menu and settings
Menu
Menu option
What it means
How to select
VT/VF rate threshold
Heart rate that must be sustained
before VT or VF is declared.
Range: 120-250 beats per minute
Defaults: VT=150 beats per
minute, VF=200 beats per minute
VT/VF response time
Elapsed time before treatment
delivered.
Range: VT=60-180 seconds,
VF=25-55 seconds
Defaults: VT=60 seconds, VF=25
seconds
MD notification option
Determines if patient receives “call
doctor” message after receiving a
treatment.
Default: Off
Can be set to repeat message
every 5 minutes for up to 1 hour
Patient sleep interval
Time the patient normally goes to
sleep and awakens.
Range: Any two times during a
24-hour clock
Default: Asleep at 00:00 (midnight)
and awake at 06:00 (6:00 a.m.)
Response time
extension
Time to be added to the response
time during the patient sleep
interval.
Range: 0-30 seconds
Default: 0
Pulse energy #1...5
Energy level of each of the five
shocks.
Range: 75-150 joules
Default: 150 joules
Post treatment
message
Message displayed after a
treatment shock.
Optional. Can be set to any
alphanumeric message you want
to display after the patient receives
treatment. For example, can be set
to display “CALL 911” or other
phone number for the patient to
call.