How to change treatment settings
Follow this procedure to change rate thresholds (in beats per minute), response
time (how many seconds until the device delivers the defibrillating shock), pulse
energy (in joules), and other treatment options.
1 With the monitor in setup mode (see page B-6), press
▲
or
▼
to
select
PATIENT
, then press
►
.
2 Press
▲
or
▼
to select
SETTINGS
, then press
►
.
3 Press
▲
or
▼
to select
TREATMENT
, then press
►
.
4 Press
▲
or
▼
until you see the setting you want to change. Then:
y
Press the response buttons so they light.
y
Press
▲
or
▼
to change the setting.
y
If the setting contains more than one option (such as VT/VF rate
threshold), press the response buttons to highlight the option you
want to change, then press
▲
or
▼
to change the setting.
y
To save the changes, press the response buttons.
y
The chart on the next page lists treatment settings, what they
mean, and how to select them.
5 Press
◄
repeatedly to return to the opening menu.