LifeVest System WCD 3100 Training Manual 

Page B-6 

ZOLL Lifecor internal use only 

How to put the monitor in setup mode 

Before doing any of the monitor setups described in this section, follow this 
procedure to put the monitor in setup mode. 

 

1    Remove and reinsert the battery. 

 

 

2    While the opening screen is displayed, hold the response buttons and 

hold 

 at the same time. Continue holding these buttons until the 

screen changes. 

This screen may be displayed for more than 10 seconds. 

 

3    Press 

 or 

 to choose a language, then press 

Note: 

The language you choose only affects the setup screens and 

will not affect the patient screens. 

 

4    The main menu displays, showing that the monitor is in setup mode. 

You may now proceed to program the monitor. 

 

5    When you are finished programming the monitor, do one of the 

following: 

y

  Navigate back to the main menu, select 

EXIT

, then press 

y

  Wherever you are in any menu, remove and reinsert the battery. 

Содержание WCD 3100

Страница 1: ...Model WCD 3100 Training Manual Prerequisite WCD 3000 training Rev B...

Страница 2: ...atient training 3 1 What to tell patients and their caregivers about the 3100 3 1 Main points to teach the patient 3 2 Summary of operating modes 3 3 Demonstrating the alarms 3 4 4 New patient setup 4...

Страница 3: ...download y how to record a manual event y how to do a pulse test What s in this manual Here s how to use this manual y What s new and different compares the 3100 to the 3000 y Patient training tells...

Страница 4: ...LifeVest System WCD 3100 Training Manual Page 1 2 ZOLL Lifecor internal use only This page intentionally left blank...

Страница 5: ...d not change Beyond the differences explained below the 3100 is identical in function and operation to the 3000 Response buttons relocated You ll notice that the response buttons are on either side of...

Страница 6: ...y Basically this is the extent of the patient s interactions more about what to teach patients is contained in section 3 y If you get the vibration alarm or the siren alarm press the response buttons...

Страница 7: ...at to do if treated Be sure to stress that patients are to interact with only the response buttons and the display There is no need for the patient to press the navigation buttons The key points to st...

Страница 8: ...you will not be able to hold the response buttons In this case and if the lethal heart rhythm continues the device will go through the treatment cycle and deliver a treatment shock y It is very import...

Страница 9: ...three operating modes Mode Message What it means What the patient needs to do Normal monitoring no alarms Monitor is operating normally Nothing Gong alarm Various messages can appear for example Pati...

Страница 10: ...er training mode View the navigation buttons from the top of the monitor with the display facing you 1 Remove and reinsert the battery 2 While the opening screen is displayed hold the response buttons...

Страница 11: ...ect TRAINING then press 6 Press to select OK then press the response buttons 7 Press the response buttons again 8 The monitor is now in training mode with these messages alternating Press the response...

Страница 12: ...g Manual Page 3 6 ZOLL Lifecor internal use only Connect patient for training mode 1 Outfit the patient with the electrode belt and garment 2 Connect the electrode belt to the monitor 3 Continue with...

Страница 13: ...red y Siren alarm sounds y Display shows that an arrhythmia has been detected and tells patient to press response buttons y Voice prompts announce to bystanders that that patient is going to be shocke...

Страница 14: ...rrect problem In this case adjust the belt so that the ECG electrodes make better contact with the skin y Press response buttons if display so states This stops the gong alarm 3 With a noise alarm whe...

Страница 15: ...Patient training ZOLL Lifecor internal use only Page 3 9 How to exit training mode 1 Remove and reinsert the battery 2 Press the response buttons as normal 3 Monitor resumes normal operation...

Страница 16: ...LifeVest System WCD 3100 Training Manual Page 3 10 ZOLL Lifecor internal use only This page intentionally left blank...

Страница 17: ...time you fit a new patient you ll need to program the device with basic patient information same as with the 3000 Programming the monitor for a new patient consists of entering the patient s name and...

Страница 18: ...LifeVest System WCD 3100 Training Manual Page 4 2 ZOLL Lifecor internal use only Menu structure...

Страница 19: ...2 While the opening screen is displayed hold the response buttons and hold at the same time Continue holding these buttons until the screen changes This screen may be displayed for more than 10 secon...

Страница 20: ...OLL Lifecor internal use only 5 When you are finished programming the monitor do one of the following to return to normal operation y Navigate back to the main menu select EXIT then press y Wherever y...

Страница 21: ...e or ID number as follows y Press and to select characters y To advance to next character press To go back press y To save the name press the response buttons 4 Enter patient s last name Follow same p...

Страница 22: ...Follow this procedure to baseline the patient 1 Prepare the patient attach the garment and electrode belt and connect the electrode cable to the monitor Have the patient sitting or lying down relaxed...

Страница 23: ...y Verify that the ECG electrodes are properly contacting the patient s skin You may need to clean the skin or clip excessive hair before fitting the patient with the electrode belt and garment y Try...

Страница 24: ...LifeVest System WCD 3100 Training Manual Page 4 8 ZOLL Lifecor internal use only This page intentionally left blank...

Страница 25: ...sembly and operation of the 3100 are the same as for the 3000 How to make a manual recording Instruct the patient as follows When you want to record a manual event hold the response buttons for three...

Страница 26: ...ecting and sending data as soon as you connect the modem to it So if you don t have the phone line connected first you will likely get an error When you get the message that the device has finished se...

Страница 27: ...nnect the modem Then reconnect the modem in the order described Connect the phone first then the monitor If you continue to get this message make sure your phone service is working There was no answer...

Страница 28: ...onitor into programming mode by doing the 3 button trick described below Then use the navigation buttons and the menu map to navigate to the pulse test We suggest that you test any monitors in storage...

Страница 29: ...buttons until the display changes This screen may be displayed for more than 10 seconds 4 Press or to choose a language then press Note The language you choose only affects the setup screens and will...

Страница 30: ...the test before completion press both response buttons y If you get an error message during the test see next page When the test is finished 1 When the test is finished one of the following messages d...

Страница 31: ...e monitor is inoperable and cannot be used Send data via modem Contact ZOLL Lifecor to report problem and arrange to have monitor replaced Battery doesn t have enough reserve power to run test Remove...

Страница 32: ...LifeVest System WCD 3100 Training Manual Page 5 8 ZOLL Lifecor internal use only This page intentionally left blank...

Страница 33: ...Quick charts ZOLL Lifecor internal use only Page A 1 Appendix A Quick charts Overview of new patient setup Program monitor for new patient Fit patient with garment and electrode belt...

Страница 34: ...LifeVest System WCD 3100 Training Manual Page A 2 ZOLL Lifecor internal use only Train patient To view patient data...

Страница 35: ...o enter setup mode To change language or second language To change patient name after initial setup To change detection settings rate threshold pulse energy response time etc To change country To chan...

Страница 36: ...t System WCD 3100 Training Manual Page A 4 ZOLL Lifecor internal use only To enter or change time zone or DST To change connectivity settings connection method modem type etc Basic troubleshooting pro...

Страница 37: ...Quick charts ZOLL Lifecor internal use only Page A 5 Troubleshooting from messages Maintenance while device is in storage...

Страница 38: ...LifeVest System WCD 3100 Training Manual Page A 6 ZOLL Lifecor internal use only This page intentionally left blank...

Страница 39: ...ix B Advanced programming About this section y This section covers programming beyond the basic monitor setup y This section applies to the WCD 3100 equipped with software version 5 1 y For basic prog...

Страница 40: ...nt s treatment settings Dialer menu Phone number dialing modes touch tone or pulse and dialing prefixes any numbers that need to be dialed in front of the phone number for dialing out when sending pat...

Страница 41: ...the monitor with the display facing you Press the arrow button that corresponds with the direction shown in the manual or on the display Once you have navigated to a menu with variables press the res...

Страница 42: ...uttons To enter a text field y Have the menu displayed with the value you want to change y Press the response buttons so they light putting the monitor in edit mode y Use or to scroll through the alph...

Страница 43: ...ers 0 through 9 If you continue pressing the up arrow you will then see symbols and a comma y To accept a character and go on to the next character press y To go back to a previous character press y T...

Страница 44: ...time Continue holding these buttons until the screen changes This screen may be displayed for more than 10 seconds 3 Press or to choose a language then press Note The language you choose only affects...

Страница 45: ...ng the setting you want to change Then y Press the response buttons so they light y Press and to select characters y To advance to next character press To go back press y To save a name change press t...

Страница 46: ...played messages and voice prompts Spanish has been approved for use in the USA y To save the change press the response buttons 5 Next you can select a secondary langauge if desired The secondary langu...

Страница 47: ...Language that monitor will use for display messages and voice prompts Select from list of languages Secondary language Second language that will be used for voice prompts to advise bystanders in addit...

Страница 48: ...ess 2 Press or to select SETTINGS then press 3 Press or to select TREATMENT then press 4 Press or until you see the setting you want to change Then y Press the response buttons so they light y Press o...

Страница 49: ...atment Default Off Can be set to repeat message every 5 minutes for up to 1 hour Patient sleep interval Time the patient normally goes to sleep and awakens Range Any two times during a 24 hour clock D...

Страница 50: ...dial prefix If using a secondary location prefix number that may be required to dial out Select any number based on what is required If any pauses are required each comma equals 1 second Hospital dial...

Страница 51: ...or assistance before changing this setting Clinical center name Name of patient s clinical center Can be any combination of letters and numbers limited to 75 characters Determined by device provider a...

Страница 52: ...r and area code The NDD number is the access code used to make a call within a country from one city to another The NDD is followed by the area city code for location being called When calling another...

Страница 53: ...Tone Hospital dial prefix Blank or determined by ZOLL Lifecor Hospital dialing number Blank or determined by ZOLL Lifecor Hospital dialer mode Tone Clinical center code Determined by ZOLL Lifecor Cli...

Страница 54: ...LifeVest System WCD 3100 Training Manual Page B 16 ZOLL Lifecor internal use only This page intentionally left blank...

Страница 55: ...quick chart A 3 D data send 5 2 data management quick chart A 2 data transfer message 5 3 data transfer problems 5 3 daylight savings time DST changing B 15 default settings B 17 demonstrating alarms...

Страница 56: ...3 phone error code 5 4 test 5 8 try again later 5 3 message post treatment changing B 12 mode defaults B 17 modem connection 5 2 error messages 5 3 modem type changing B 16 monitor setup for new pati...

Страница 57: ...B 12 setup quick charts A 1 setup for a new patient 4 1 setup mode 4 3 B 7 quick chart A 3 shock 3 2 siren alarm 3 2 storage quick chart A 5 T teaching patient 3 1 test plug 5 5 testing monitor 5 5 pr...

Страница 58: ...LifeVest System WCD 3100 Training Manual Index 4 ZOLL Lifecor internal use This page intentionally left blank...

Отзывы: