Introduction
10
CAUTION! START-UP KIT LIFETIME IS SEVEN (7) DAYS.
The designed operating lifetime for start-up kit components is seven (7) days
of continuous operation. If a patient must be treated for a longer period, a new
start-up kit must be installed in the System. Failure to adhere to this time limit
may cause injury to the patient.
Data Memory
The System is capable of continuously recording patient temperature and system
activity for up to 21 days. This stored data can be downloaded to an attached
computer over a serial interface using optional software furnished by ZOLL.
Saline Circuit Diagram
A flow diagram is printed on the inside of the top cover. Use this diagram to
assure that the start-up kit has been installed correctly.
Figure 2-8. Flow Diagram.
600479-001 Rev 2 22/147
Содержание CoolGard 3000
Страница 1: ...Operation Manual 600479 001 Rev 2 1 147...
Страница 4: ...Safety Information 1 Safety Information 600479 001 Rev 2 4 147...
Страница 13: ...Introduction 1 Introduction 600479 001 Rev 2 13 147...
Страница 25: ...Receiving Inspection Assembly 1 Receiving Inspection Assembly 600479 001 Rev 2 25 147...
Страница 32: ...Operation 1 Operation 600479 001 Rev 2 32 147...
Страница 95: ...Accesories HMIA 1 Hospital Monitor Interface Accessory HMIA 600479 001 Rev 2 95 147...
Страница 109: ...TempTrend CSV 1 TempTrend CSV 600479 001 Rev 2 109 147...
Страница 120: ...Alarms Corrective Actions 1 Alarms Corrective Actions 600479 001 Rev 2 120 147...
Страница 125: ...Troubleshooting 1 Troubleshooting 600479 001 Rev 2 125 147...
Страница 131: ...Maintenance 1 Maintenance 600479 001 Rev 2 131 147...
Страница 140: ...Warranty Service 1 Warranty Service 600479 001 Rev 2 140 147...
Страница 145: ...Specifications 1 Specifications 600479 001 Rev 2 145 147...