Operation
49
Figure 4-60. Location of the Serial Interface Connector.
11. Start
the
TempTrend
CSV
program on the laptop computer. Refer to
Chapter 5 - TempTrend CSV.
12.
If a problem occurs during downloading, the System screen will display
the message “Download error. Please check external computer.” Check
the serial cable connections and the operation of the laptop computer
and choose “
Try again
.” If repeated failures occur, choose “
Cancel
.”
Repeated failures indicate a problem with the
TempTrend
program or the
System; contact your ZOLL representative for assistance.
Figure 4-61. Download Error Message.
13.
When the data download is complete, the System screen briefly displays
the message “
Download Complete
.” At this time, the patient trend data
saved from the improper shutdown is deleted.
600479-001 Rev 2 80/147
Содержание CoolGard 3000
Страница 1: ...Operation Manual 600479 001 Rev 2 1 147...
Страница 4: ...Safety Information 1 Safety Information 600479 001 Rev 2 4 147...
Страница 13: ...Introduction 1 Introduction 600479 001 Rev 2 13 147...
Страница 25: ...Receiving Inspection Assembly 1 Receiving Inspection Assembly 600479 001 Rev 2 25 147...
Страница 32: ...Operation 1 Operation 600479 001 Rev 2 32 147...
Страница 95: ...Accesories HMIA 1 Hospital Monitor Interface Accessory HMIA 600479 001 Rev 2 95 147...
Страница 109: ...TempTrend CSV 1 TempTrend CSV 600479 001 Rev 2 109 147...
Страница 120: ...Alarms Corrective Actions 1 Alarms Corrective Actions 600479 001 Rev 2 120 147...
Страница 125: ...Troubleshooting 1 Troubleshooting 600479 001 Rev 2 125 147...
Страница 131: ...Maintenance 1 Maintenance 600479 001 Rev 2 131 147...
Страница 140: ...Warranty Service 1 Warranty Service 600479 001 Rev 2 140 147...
Страница 145: ...Specifications 1 Specifications 600479 001 Rev 2 145 147...