TROUBLESHOOTING GUIDES
III-3
Fault Codes Table
When the FAULT indicator light illuminates, press and hold the
TEST
button. The
illumination of the following indicator lights together represent a specific defect
identified under the
Fault Code
column.
Charger On
Batt. Ready
Fault
Testing
Fault Code
ON
OFF
OFF
OFF
Constant
Current Voltage
ON
ON
OFF
OFF
Discharge
ON
ON
ON
OFF
Discharge Time
ON
ON
ON
ON
OverCharge
ON
ON
OFF
ON
Temperature
ON
OFF
ON
ON
Low Battery
Voltage
ON
OFF
ON
OFF
High Constant
Current
ON
OFF
OFF
ON
Shutdown
OFF
ON
ON
ON
High Battery
Voltage
OFF
OFF
ON
ON
Battery
Capacity
OFF
OFF
OFF
ON
Charger
OFF
ON
OFF
OFF
Constant
Voltage
OFF
ON
OFF
ON
Vcc Voltage
OFF
ON
ON
OFF
Low Power
Supply Voltage
OFF
OFF
ON
OFF
Low Constant
Current
NOTE:
A battery that has failed a capacity test will display the
TESTING
&
FAULT
LED's simultaneously. When the
TEST
button is pressed and held the LED pattern
will not change. This is a normal operating condition and does not reflect a fault with
the unit.
Содержание Base Powercharger 4x4
Страница 1: ...9650 0072 01 Rev E ZOLL Base Powercharger 4x4 Service Manual...
Страница 8: ...GENERAL INFORMATION I 7 SEE FILE PG7_C PM5 FOR THIS PAGE AND ALL OF THE SYMBOLS...
Страница 11: ...GENERAL INFORMATION I 10...
Страница 13: ...CHECKOUT PROCEDURES 9650 0072 II 2 Rev D...
Страница 39: ...SCHEMATIC DRAWINGS V 2...
Страница 49: ...COMPONENT LAYOUT DRAWINGS VI 2...
Страница 53: ...DISASSEMBLY PROCEDURES VII 2...
Страница 55: ...DISASSEMBLY PROCEDURES VII 4...
Страница 57: ...DISASSEMBLY PROCEDURES VII 6...
Страница 58: ...DISASSEMBLY PROCEDURES VII 7...
Страница 60: ...DISASSEMBLY PROCEDURES VII 9...
Страница 62: ...DISASSEMBLY PROCEDURES VII 11...
Страница 63: ...DISASSEMBLY PROCEDURES VII 12...
Страница 65: ...DISASSEMBLY PROCEDURES VII 14...
Страница 67: ...DISASSEMBLY PROCEDURES VII 16...
Страница 71: ...REPLACEMENT PARTS VIII 4...
Страница 73: ...A 2 Base PowerCharger4x4 Operator s Guide...
Страница 75: ...B2 Base PowerCharger4x4 Operator s Guide...