TROUBLESHOOTING GUIDES
III-2
TROUBLESHOOTING TABLE
Symptoms
TroubleShooting Method and Possible
Causes
POWER
light does not illuminate.
•
Check that the unit is plugged into an
appropriate power source.
•
Check for defective power cord.
•
Plug unit into another power source.
•
Check for blown fuse. Disconnect power
cord and open fuse holder to check fuses.
CHARGER ON
indicator light does not
illuminate when battery is inserted into
compartment.
•
Check that the battery is properly seated
in battery compartment.
•
Check that battery contacts are clean and
not damaged.
•
Check that unit is plugged into an
appropriate power source.
•
Test the battery compartment for
functionality.
•
Place another battery in the compartment.
TESTING
light does not illuminate when
TEST
button is pressed.
•
Check that the battery is properly seated
in compartment.
•
Check that unit is plugged into an
appropriate power source.
•
Remove the battery and press the
TEST
button to perform a compartment self-test.
If it does not self-test the button is
defective.
Battery fails test cycle. (
FAULT
light
illuminates with
TESTING
light).
•
Verify that the battery compartment is fully
operational with no defects.
•
Press and hold the
TEST
button to
determine the fault code. (See
Fault
Codes Table
on Page III-3).
•
If all
FAULT
LED's are on most likely
there is a failure in the common circuitry.
When the unit does not follow the Power-Up sequence explained elsewhere in this
manual or if a fault light comes on at any time follow the steps listed below to identify
the fault.
Note:
The unit can be run through it 's self-test at any time, as long as there is no
battery installed, by pressing the
TEST
button. When the self-test is invoked on a
single battery compartment the unit will respond by lighting all the LED's associated
with the compartment for a few seconds. This will be followed by a beep and then the
TESTING
and
BATT. READY
LED's will light for a couple of seconds. Then all LED's
extinguish.
Faults that may endanger the reliable charging of a battery or the ability to
charge/test successfully are latching faults that can only be cleared by power cycling
after the fault is fixed. All other faults are self-clearing as soon as the fault condition is
removed.
Содержание Base Powercharger 4x4
Страница 1: ...9650 0072 01 Rev E ZOLL Base Powercharger 4x4 Service Manual...
Страница 8: ...GENERAL INFORMATION I 7 SEE FILE PG7_C PM5 FOR THIS PAGE AND ALL OF THE SYMBOLS...
Страница 11: ...GENERAL INFORMATION I 10...
Страница 13: ...CHECKOUT PROCEDURES 9650 0072 II 2 Rev D...
Страница 39: ...SCHEMATIC DRAWINGS V 2...
Страница 49: ...COMPONENT LAYOUT DRAWINGS VI 2...
Страница 53: ...DISASSEMBLY PROCEDURES VII 2...
Страница 55: ...DISASSEMBLY PROCEDURES VII 4...
Страница 57: ...DISASSEMBLY PROCEDURES VII 6...
Страница 58: ...DISASSEMBLY PROCEDURES VII 7...
Страница 60: ...DISASSEMBLY PROCEDURES VII 9...
Страница 62: ...DISASSEMBLY PROCEDURES VII 11...
Страница 63: ...DISASSEMBLY PROCEDURES VII 12...
Страница 65: ...DISASSEMBLY PROCEDURES VII 14...
Страница 67: ...DISASSEMBLY PROCEDURES VII 16...
Страница 71: ...REPLACEMENT PARTS VIII 4...
Страница 73: ...A 2 Base PowerCharger4x4 Operator s Guide...
Страница 75: ...B2 Base PowerCharger4x4 Operator s Guide...