906-0731-01-05 Rev. C
ZOLL Ventilator Operator’s Guide
A-1
Appendix A
Specifications
General
Parameter
Operating Range
Operating Modes
EMV+
®
and Eagle II
™
: AC,SIMV, CPAP with and without Pressure Support,
BL Modes, and Leak Compensation for Active and Noninvasive ventilation.
AEV
®
: AC, CPAP (with and without Pressure Support), BL Modes, and Leak
Compensation for Noninvasive ventilation.)
Breath Target
Volume or Pressure
Flow Rate
0 to 100 LPM at 40 cm H
2
O
Breath Rate
1 to 80 BPM ±1 BPM over the interval
Setting Resolution: 1 BPM
Measurement: 1 to 90 BPM ± 1 BPM over the interval
Inspiratory Time (Ti)
Setting: 0 to 3 ± s for I:E ratio > 1:1.0, 0 to 5 ± s for E:I ratio <= 1.0:1
Setting Resolution: 0.05 s
Tidal Volume
Setting: 50 to 2000 ml ATPD ± (5 ml +10% setting)
Setting Resolution: 10 ml
Measurement: 0 to 9999 ml ATPD ± (5 ml +10% setting)
FIO
2
21 to 100% (± 3% of full scale ± 10% of setting)
PEEP/EPAP
Setting: 0 to 30 cm H
2
O ± (2 cm H
2
O + 8% of reading)
Setting Resolution: 10 ml
Measurement: 0 to 30 cm H
2
O ± (2 cm H
2
O + 8% of reading)
Peak Inspiratory Pressure
(PIP)
Setting: 10 to 80 cm H
2
O (± 2 cm H
2
O ± 8% of setting)
Setting Resolution: 1 cm H
2
O
Measurement: 0 to 99 cm H
2
O ± (2 cm H
2
O + 8% of reading)