A
LARMS
5-28
www.zoll.com
906-0731-01-05 Rev. C
PEEP Setting Conflict
Popup triggers when the user attempts to set the PEEP setting
≤
5 cm H
2
O below the PIP High pressure
limit.
Message: Cannot adjust PEEP target to within 5 of PIP High Limits
PEEP Setting Conflict
Popup triggers when the user attempts to configure the ventilator so that the PEEP plus the pressure sup-
port (PS) are > the PIP High pressure limit.
Message: PEEP + PS cannot be greater than PIP High Limit
PEEP Backup Setting Conflict
Popup triggers when the user attempts to set the PEEP setting
≤
5 cm H
2
O below the Apnea Backup PIP
pressure during CPAP or BL mode ventilation.
Message: Cannot adjust PEEP target to within 5 of backup PIP target
PEEP Setting Conflict
Popup triggers when the user attempts to set the PEEP
≤
5 cm H
2
O below PIP pressure.
Message: Cannot adjust PEEP target to within 5 of PIP target
PEEP+PS Setting Conflict
Popup triggers when the user attempts to set a combination of PEEP and PS that is <3 cm H
2
O.
Message: Cannot adjust PEEP+PS below 3
High Pressure Target Setting
Popup triggers when the user attempts to set the PIP pressure >60 cm H
2
O. To do this, the user must press
the Accept button and then continue to set a value >60 cm H
2
O followed by Accept again to confirm the set-
ting change.
Message: Confirmation required -- press accept key to exceed 60 cmH2O
PIP Setting Conflict
Popup triggers when the user attempts to set the PIP target
≤
5 of PEEP pressure.
Message: Cannot adjust PIP target to less than 5 more than PEEP
PIP Setting Conflict
Popup triggers when the user attempts to set the PIP > the PIP High pressure limit.
Message: Cannot adjust PIP target higher than PIP High Limit
BPM Limit Conflict
Popup triggers when the user attempts to set the PIP High pressure limit < the PIP Low pressure limit.
Message: Cannot adjust high limit lower than low limit
Low Breath Rate Setting
Popup triggers when the user attempts to set the BPM < 6 bpm. Doing this could, in effect, disable the
alarm for some patients. To do this, the user must press the Accept button and then continue to set a value
<6 bpm followed by Accept again to confirm the setting change.
Message: Confirmation required -- press accept key for values below 6 BPM
BPM Limit Conflict
Popup triggers when the user attempts to set the BPM Low limit > the BPM High limit.
Message: Cannot adjust low limit higher than high limit
Vt Limit Conflict
Popup triggers when the user attempts to set the Vt high limit < the Vt low limit.
Message: Cannot adjust high limit lower than low limit