C
HAPTER
1 M
AINTENANCE
T
ESTS
1–8
www.zoll.com
9650-0308-01 Rev. A
Put the puck down and turn off the AED Plus unit.
Disconnect the universal adapter cable from the AED Plus unit.
2.8
Wait 10 seconds and then turn on
the AED Plus unit.
!
PLUG IN CABLE
This prompt is optional:
—
STAY CALM
!
CHECK RESPONSIVENESS
(or
CHECK PATIENT
)
This prompt is optional:
—
CALL FOR HELP
These three prompts are optional:
—
OPEN AIRWAY
—
CHECK BREATHING
—
GIVE TWO BREATHS
!
CHECK CIRCULATION
(or
CHECK PULSE
)
!
PLUG IN CABLE
!
!
2.9
Check the expiration date on the
previously connected defibrillation
electrodes.
If the electrodes have not expired,
reconnect them to the unit;
otherwise, connect a fresh
package of defibrillation
electrodes.
!
ADULT PADS
(or
PEDIATRIC PADS
)
!
!
Turn off the AED Plus unit.
2.10
After 90 seconds, check the
Status indicator on the unit.
•
The Status indicator shows a green
check.
!
!
Test
Action
Expected Result(s)
Pass
Fail