Preventive Maintenance Test
9650-0308-01 Rev. A
ZOLL AED Plus Service Manual
1–5
Test Procedure
Test
Action
Expected Result(s)
Pass
Fail
2.1
Turn on the AED Plus unit.
!
UNIT OK
!
ADULT PADS
!
DON’T TOUCH PATIENT,
ANALYZING
!
TREATMENT ADVISED
(or
SHOCK ADVISED
)
!
DON’T TOUCH PATIENT.
!
PRESS TREATMENT BUTTON
(or
PRESS SHOCK BUTTON
)
!
!
2.2
Press the
Treatment
button
(first shock).
•
Analyzer indicates a defibrillator
discharge at the configured
first-shock energy level (120, 150
or 200 joules ±15%).
!
TREATMENT DELIVERED
(or
SHOCK DELIVERED
)
•
An updated shock count appears
on the display (
SHOCKS: 1
).
!
DON’T TOUCH PATIENT,
ANALYZING
!
TREATMENT ADVISED
(or
SHOCK ADVISED
)
!
DON’T TOUCH PATIENT.
!
PRESS TREATMENT BUTTON
(or
PRESS SHOCK BUTTON
)
!
!