ZOLL AED PLUS Administrator’s Guide
29
Guidance and Manufacturer’s Declaration - Electromagnetic
Emissions
Table 8: EMC Specifications
The ZOLL AED Plus is intended for use in the electromagnetic environment specified below.
The customer or the user of the ZOLL AED Plus should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment -
guidance
RF emissions CISPR11
Group 1
The ZOLL AED Plus uses RF
energy only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions CISPR 11
Class B
Harmonic Emission
IEC 61000 3-2
Not applicable
Voltage Fluctuations/Flicker
Emission
IEC 61000 3-3
Not applicable
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and
put into service according to EMC information provided in this document.
The ZOLL AED Plus is intended for use in the electromagnetic environment specified below.
The customer or the user of the ZOLL AED Plus should assure that it is used in such an
environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Not applicable
± 1 kV I/O
Surge
IEC 61000-4-5
± 1 kV differential
mode
+/- 2 kV common
mode
Not applicable
Not applicable