ZOLL AED PLUS Administrator’s Guide
11
Using Electrodes
The ZOLL AED Plus supports both adult and pediatric electrode pads. The device automatically
adjusts defibrillation energy to adult or pediatric levels depending on the type of electrodes connected
to it. Make sure that the electrode pads used are appropriate for the patient.
WARNING! DO NOT use adult electrode pads or CPR-D padz on patients under 8 years of age.
The ZOLL AED Plus uses electrode packs that are connected to the unit by a cable. The package
contains electrodes that you attach to the patient.
•
Make sure to install a new package of electrodes and connect the electrode cable to the unit after
each use, to prepare for future emergencies.
•
Check the electrode expiration date regularly to ensure that electrodes are fresh and ready to use
in an emergency situation.
•
Replace electrodes, if expired
•
After the ZOLL AED Plus is powered on and completes its self-test, the unit will issue an “Adult
Pads” voice prompt or a “Pediatric Pads” voice prompt to indicate the type of electrodes connected
to the unit. Verify that the connected electrode pads are appropriate for the patient being treated.
Connect alternate electrode pads if necessary.
If the electrodes are not attached properly, the unit issues the
CHECK ELECTRODE PADS
or the
ATTACH ELECTRODE PADS
voice prompt during operation. If the electrode cable is not properly
attached to the unit, the unit issues the
PLUG IN CABLE
prompt. Make certain to connect the electrode
cable to the ZOLL AED Plus unit and attach the electrodes to the patient properly.
WARNING! Electrodes must be pre-attached to the device. Keep the electrode cable connected
to the ZOLL AED Plus unit at all times.
The electrode package may include:
• Scissors to cut clothing or chest hair.
• Razor to remove excessive hair at the electrode application site, if necessary.
• Small towel to make sure that the patient’s skin is dry.
• Gloves.
• Barrier Mask.
NOTE Electrodes contain no hazardous materials and may be disposed of in general trash unless contaminated
with pathogens. If contaminated, appropriate precautions should be used in their disposal.
WARNING!
DO NOT reuse electrodes.