4.2 

Measurement Steps 

After opening the TSic Recorder™ software, follow the steps described below to acquire measurement data. 

Step1

:   Connect at least one of the four temperature sensors to the TSic LABkit™ Adapter and select the 

connected temperature channels in the main window of the TSic Recorder™ software. 

 

Figure 4.1  – TSic Recorder™ Software Main Window 

Step 2

:  Correct settings if necessary using the Settings dialog window described in section 4.4. Then 

click OK to exit the Settings dialog window. For all enabled temperature channels, the actual 
measurement values are displayed in the main window (see Figure 4.1 above). The TSic 
LABkit™ system is now ready to read the sensors’ measured temperature. 

Important: 

The software default configuration after installation is for the TSic™ 306. If other TSic™ types are 

connected, see section 4.4.2 below for required sensor type configuration using the Settings dialog window. 

The Settings dialog window can also be used to change settings such as temperature display refresh time and 
the interval of the displayed mean value.  

See section 4.3 for more information on the main window settings. 

Troubleshooting Note: If measurement data is missing, check the connections (USB connector, TSic LABkit™ 
Adapter and sensors) and ensure that the correct USB Port has been selected. You can select the USB/COM 
port in the Settings dialog window as described in section 4.4, Figure 4.2. 

 

Step 3

:  To record temperature measurement values in a text file, click the Start button in the main window. 

Click the Stop button to end recording.  

The default directory and filename are C:TSIC-LABKIT\DATA\tempdata.csv. To change the directory or 
filename, see section 4.4.7.  

The information to be recorded with the sensor data in the text file can be selected on the main window. By 
default, the text file includes the measurement number, date, and time columns and the Text 1-3 file heading 
text in addition to the “Actual” (T1-4) and “Raw Value” (R1-4) data for each enabled sensor.  

Содержание TSic LABkit

Страница 1: ...WARE 10 5 1 USB HARDWARE CONFIGURATION 10 5 2 MICROCONTROLLERS IN CIRCUIT PROGRAMMING 10 6 SENSORS FOR THE TSIC LABKIT TM SENSOR CONNECTION CABLE AND CONNECTOR 11 6 1 STANDARD CABLES 11 6 2 REQUIREMEN...

Страница 2: ...wner 1 TSic LABkit Contents a CD ROM with the TSic Recorder Software for Microsoft Windows b TSic LABkit TM Adapter for USB c Four TSic TM 306 Sensors with Cable Connector d USB Cable Figure 1 1 LABki...

Страница 3: ...rface USB Adapter Dimensions 80 mm x 40 mm x 23 mm EMC Noise Emission Noise Immunity 89 336 EEC European Harmonized Standards EN 61000 6 3 2001 EN 61000 6 2 2001 Note The technical data provided above...

Страница 4: ...soft Excel 4 1 System Requirements and Software Installation System Requirements 5x86 compatible PC 32 MB RAM Hard drive with 20MB free space USB port Microsoft W98 ME 2000 XP Internet Explorer 6 0 re...

Страница 5: ...after installation is for the TSic 306 If other TSic types are connected see section 4 4 2 below for required sensor type configuration using the Settings dialog window The Settings dialog window can...

Страница 6: ...urements is selected in the Settings dialog window See section 4 4 The Raw Value readings are the decimal values output by the TSic sensors Conversion into C or F values is based on the conversion tab...

Страница 7: ...in the main window To close the Settings dialog window and apply any new settings click OK Figure 4 2 TSic Recorder Settings Dialog Window Figure 4 3 USB COM Port Setting 4 4 1 USB COM Port Settings I...

Страница 8: ...figure the correct measurement range Note In addition to the standard sensor products offered by IST customized OEM sensor types are also available from IST and can be manually configured as specified...

Страница 9: ...ta is recorded enter the new directory and filename which must be write accessible in the Path dialog box Default directory and filename C TSIC LABKIT DATA tempdata csv Change Filename at 12 00 a m Se...

Страница 10: ...protection 5 1 USB Hardware Configuration The jumpers on the PCB control the following settings Important Connect either JP1 or JP8 not both at the same time Pin Connections Name Settings JP1 1 2 GND...

Страница 11: ...imum temperature 80 C Blocking capacitor 100nF connected close to the sensor between VDD and GND Note If the standard cable requires an extention add it on the connector side of the cable to minimize...

Страница 12: ...sor between VDD and GND 3 pole connector Schukat Electronics part no NSG25 3 crimp contact NSK25 0 7 Electromagnetic Compatibility EMC Standard 89 336 EU The TSic LABKIT USB Adapter meets the electrom...

Страница 13: ...sent continuously and separated by semicolon The character and carriage return CR mark the end of the transmission Typical data line structure example 08DA 08C2 08CC 08CA 2332 CR Characters Definition...

Страница 14: ...0 C 0x7FF 07FF XXXX XXXX XXXX CCCC CR Not valid 0x800 to 0xFFF UUUU XXXX XXXX XXXX CCCC CR No response Timeout no sensor FFFF XXXX XXXX XXXX CCCC CR XXXX Other channels CCCC Value for check sum UUUU V...

Страница 15: ...C accuracy 0 1 C for 5 C to 45 C Test Messung 1 Text2 Text3 Sensoren 4x TSic 306 Lauf Nr Date Time Sensor 1 Sensor 2 Sensor 3 Sensor 4 2897 08 08 2005 17 39 39 24 65 24 74 24 74 24 55 2898 08 08 2005...

Страница 16: ...23 02 22 23 58 22 24 54 22 25 51 22 26 47 22 27 43 22 28 39 T1 T2 T3 T4 57 57 5 58 58 5 59 59 5 60 60 5 61 61 5 62 62 5 63 63 5 22 14 42 22 14 52 22 15 02 22 15 12 22 15 22 T1 T2 T3 T4 Example 1 Proc...

Страница 17: ...in the accuracy range The chart shows the higher resolution of TSic 506F which is shown as sensor T4 Figure 8 6 Surface Temperature Measurement Before and After Touching the Sensors Number Date Time...

Страница 18: ...licensee or third party for any damages including but not limited to personal injury property damage loss of profits loss of use interruption of business or indirect special incidental or consequentia...

Отзывы: