Manufacturer’s declaration of
Electromagnetic Compatibility
15
46
The main features of the en
Puls
Version 2.0 are as follows: interference-free delivery of shockwaves, interference-free
control of all functions. Uninterrupted operation is not required with the use intended.
Guidelines and manufacturer's declaration – electromagnetic interference immunity
The device en
Puls
Version 2.0 is intended for operation in the electromagnetic environment specified below. The customer
or user of the en
Puls
Version 2.0 should ensure that it is used in such an environment.
Interference
immunity tests
IEC 60601-test level
Compliance level
Electromagnetic environment - guidelines
Conducted RF
disturbance variables
according to IEC
61000-4-6
Radiated RF
disturbance variables
according to IEC
61000-4-3
3 V
effektive value
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
effektive value
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Portable and mobile radio equipment should
not be used any closer to the en
Puls
Version
2.0, including cables, than the recommended
separation distance calculated from the
equation applicable to the transmission
frequency
.
Recommended separation distance:
d= 1.2 √P
d= 0.35 √P for 80 MHz to 800 MHz
d= 0,7 √P for 800 MHz to 2.5 GHz
Where P is the rated power of the transmitter
in Watts (W) according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
According to an investigation in situ
a
, the field
strength of stationary radio transmitters should
be less than the compliance level at all
frequencies.
Interference may occur in the vicinity of
equipment which is marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz the higher frequency range is applicable.
NOTE 2 These guidelines may not be applicable in all cases. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.