1.5. Applicator heads
1
12
There are 3 different applicator heads available for treatment.
Changing
applicator heads
To change the various applicator heads, hold the handpiece in one hand
and unscrew the applicator head from the handpiece with the other hand
(anticlockwise). Screw the required head tightly onto the handpiece
(clockwise), until the black outside ring of the applicator head rests on the
handpiece (there should no longer be any thread visible).
Note:
Applicator heads are expendable parts and must be replaced after a
certain period of use.
Minor / slight deformation or shortening of the rear impact dome does not
affect functionality.
In cases of greater deformation or stronger shortening of the rear impact
dome the applicator head must be replaced.
A test template is supplied with the device that enables the user to test if
the wear limit has been reached
(see diagram).
Air gap
applicator OK
Test template
Test template
Template makes contact or air gap at the tip
Wear limit has been reached