2.1 INITIAL INSPECTION
Unpack the
A.T.S.
1200 Tourniquet upon receipt and inspect
the unit for any obvious damage that may have occurred
during shipment. We recommend that this inspection be
performed by a qualified biomedical engineer or other person
thoroughly familiar with electronic medical devices. If the
unit is damaged, notify the carrier and your Zimmer
representative immediately. If the initial inspection results
are satisfactory, a functional and calibration check should be
performed after a 24-hour charge. The attention label
covering the ON/STANDBY button can be removed and
discarded after the 24-hour charge.
2.2 CONTROLS, INDICATORS,
AND CONNECTORS
Refer to Figure 1 and 2 in the back of the manual for the
locations of the unit’s controls, indicators, and connectors.
1.
ON/STANDBY Button
Turns the unit ON or sets the unit to STANDBY. This
button will not set the unit to STANDBY when the cuff
pressure is at a non-zero value.
Ensure both cuffs are
fully deflated and have been removed from the
patient prior to setting the unit to STANDBY.
NOTE: During STANDBY, the power to the
A.T.S.
1200 instrument and all instrument functions
(i.e. inflation, deflation, etc.) are OFF but power
continues to supply the battery charging circuitry
anytime ~ (AC) power (Mains) is present.
2.
ROTARY knob
Changes the value of set time or default time and set
pressure or default. Turn knob clockwise to increase the
value; turn knob counterclockwise to decrease the value.
3.
PRESSURE button
Press to verify or modify set pressure.
4.
TIME button
Press to verify or modify set time.
5.
MAIN CUFF INFLATE button
Inflation of the Main cuff is initiated by depressing the
red
INFLATE button.
6.
MAIN CUFF DEFLATE button
Deflation of the Main cuff is initiated by depressing the
red
DEFLATE button. For greater safety, the DEFLATE
button has a delay and, therefore, must be held for
approximately 2 seconds before the unit will allow a cuff
to deflate.
7.
SECOND CUFF INFLATE button
Inflation of the Second cuff is initiated by depressing the
blue
INFLATE button.
8.
SECOND CUFF DEFLATE button
Deflation of the Second cuff is initiated by depressing
the
blue
DEFLATE button. For greater safety, the
DEFLATE button has a delay and, therefore, must be
held for approximately 2 seconds before the unit will
allow a cuff to deflate.
9.
ALARM SILENCE button
The ALARM SILENCE button will silence most audible
alarms for 30 seconds after the button is pressed. When
an alarm sounds because of an internal hardware
malfunction, the alarm cannot be silenced.
NOTE:
The alarm messages will continue to flash on
the displays until the alarm condition is corrected.
10.
AC MAINS indicator light
The AC MAINS indicator light indicates that the unit is
plugged in and is being powered by AC Mains. This is
the normal means of operation (battery power is only
intended for emergency power loss or patient transport).
11.
BATTERY indicator light
The BATTERY indicator light indicates that the unit
is operating on backup battery. The light will flash
continuously while the unit is running on battery
backup power.
12.
PRESSURE display
During normal operation with no buttons being pressed,
the PRESSURE display will show the monitored cuff(s)
pressure. At other times, depending on alarm conditions
and buttons pressed, this display may communicate
other information such as alarm messages, set pressure,
or default set pressure.
13.
TIME display
During normal operation with no buttons being pressed,
the TIME display will show elapsed inflation time of the
first cuff that was inflated in 1-minute increments. At
other times, depending on alarm conditions and buttons
pressed, this display may communicate other information
such as alarm messages, set time, or default set time.
NOTE:
The elapsed inflation time can be “zeroed” at
any point in the procedure by pressing the TIME and
PRESSURE buttons simultaneously.
INSTALLATION
AND OPERATING
INSTRUCTIONS
SECTION 2.0
A.T.S. 1200 TOURNIQUET SYSTEM
7