Table 3.2 Troubleshooting
SYMPTOM
POSSIBLE CAUSES
1. Main cuff or Second cuff will not inflate.
a) Membrane Panel not properly plugged into P8.
b) Tubing inside unit may be pinched or improperly connect-
ed.
c) Deflate valve is stuck open.
d) Pump not properly plugged into P5.
e) Pump’s electrical harness damaged.
f) INFLATE button not working.
g) Valve’s electrical harness damaged (P6).
h) Defective valve driver circuitry.
2. Main cuff or Second cuff will not deflate.
a) Membrane Panel not properly plugged into P8.
b) DEFLATE button not pressed long enough
(at least 2 seconds).
c) Deflate valve is stuck shut.
d) DEFLATE button not working.
e) Valve’s electrical harness damaged (P6).
f) Defective valve driver circuitry.
3. No
green
AC MAINS indicator light.
a) Unit not plugged into wall outlet.
b) No Power at wall outlet.
c) Mains AC harness not properly plugged into P1.
d) Blown fuse(s).
e) Membrane Panel not properly plugged into P8.
f) Defective AC MAINS indicator.
g) Defective AC MAINS indicator circuitry.
4. No flashing
orange
BATTERY indicator light.
a) Unit running on AC.
b) Membrane Panel not properly plugged into P8.
c) Defective BATTERY indicator.
d) Defective BATTERY indicator circuitry.
5. ALARM SILENCE button not working.
a) Membrane panel not properly plugged into P8.
b) Non-silenceable alarm (System Failure).
c) ALARM SILENCE button defective.
d) Defective alarm silence circuitry.
6. ALARM indicator light not working.
a) Membrane panel not properly plugged into P8.
b) Defective ALARM indicator.
c) Defective ALARM indicator circuitry.
7. No cuff pressure reading.
a) Transducer amplifier not working.
b) Internal tubing kinked.
c) Transducer tubing on incorrect transducer.
8. Pump will not stop running.
a) Leak in internal hose or connector.
b) Internal tubing kinked.
c) Transducer(s) not working.
d) Transducer tubing on incorrect transducer.
9. BATTERY FAIL alarm/message.
a) Blown battery fuse (board mounted F1).
b) Broken Battery wire harness.
c) Dead or depleted battery.
19
A.T.S. 1200 TOURNIQUET SYSTEM
Содержание 60-1200-101-00
Страница 1: ...Operator Service Manual A T S 1200 TOURNIQUET SYSTEM REF 60 1200 101 00 ...
Страница 26: ...R Power Cord S Cord Retaining Clip T Factory Test Port U Pole Clamp V Mains Fuse Block 24 2 ...
Страница 30: ...28 9 Control Board Layout ...
Страница 32: ...30 10 ...
Страница 33: ...SCHEMATICS ...