7 |
iASSIST 2-Pod Version
Surgical Technique
Figure 2.6
Pod Status Bar
Idle
Cut Guide
Not Connected
Navigation Pod
Connected to
Femur Reference
Buttons, Icon and Pod
Status Bar
(cont.)
Patient Side Icon
The patient side is indicated at the top of the system
controller computer screen. When the patient side is
set, both the icon and letter are green on the selected
side. If the patient side is not determined the icon is
gray with a “?” sign.
Pod Status Bar
The “Pod Status” bar is displayed at the bottom of the
system controller computer screen. The bar displays
the status of each pod connected to the system
controller computer showing its user interface.
A pod that has not joined the network will be displayed
in red.
A pod that has joined the network but has not been
connected to an instrument will be displayed in yellow.
A pod that has joined the network and is connected to
an instrument will be displayed in grey. The name of
the instrument on which the pod is connected is also
displayed.
Issues with the pod’s battery life, temperature or
communication will be displayed on the “Pod Status”
bar.